5 Installation
5.1 Installing the pump
Installation location
● Indoors, protected from dust deposits
● Outdoors, protected from direct weather influences
Installation altitude
max. 2000 m
Orientation
as level as possible; max. permissible angle of inclination: ±10°
Ambient temperature
+12°C to +40°C
Relative air humidity
max. 85%
Tbl. 5:
Ambient conditions
max. 10
°
> 35 mm
Fig. 6:
Minimum clearances to side boundaries and permissible inclination
Notes on installing the pump
1. Fill with operating fluid prior to first commissioning.
– Quantity and type according to rating plate.
2. Place the pump on a flat, horizontal surface, to safeguard the operating fluid supply.
3. Screw the base plate of the pump to the mounting surface if necessary.
4. When installing the pump in a closed housing, ensure adequate air circulation.
5. Keep the sight glass and gas ballast valve visible and freely accessible.
6. Keep the voltage and frequency specifications on the motor rating plate visible and freely accessi-
ble.
Installation
17/66