BodiLink Head Support Operation Manual
9
INSTALLATION - ACTA-RELIEF WITH COMPASS POWER HARDWARE
For Acta-Relief and Acta-Relief LTS using Compass Power Mount Hardware.
1
If the head support hardware stem (C)
is inserted in the mounting block (B), turn the
knob (A) counterclockwise to release it. Remove the hardware stem; set aside.
2
Using the 5mm hex key, loosen the screws (D) to release the washers (E), the spacer
plate (F), and the double nuts
(G) from the mounting block (B).
WARNING:
Using the spacer plate (F) may affect clearance in certain recline situations.
Note: If the optional spacer plate F is not used, replace screws (D) with the provided shorter screws*.
*M6 X 1 X 10mm socket head cap screws
3
Align the mounting block (B) with the desired set of head support holes on the power
mount bracket (H)
and with the double nuts (G).
4
Replace the screws (D) and washers (E) and tighten.
5
Insert the hardware stem (C)
into mounting block (B)
and turn knob (A) clockwise to
secure.
IMPORTANT!
Make sure that at least ½” of the hardware stem (C) is completely
through the bottom of mounting block (B) at all times. See the warning in the
“Important
Safety Information” section in this manual.
Note: See SUPPORT PAD installation, positioning and adjustment instructions in this manual to make final adjustments.
1
A
G
F
B
C
D
E
H
G
D
E
B