background image

2

•  Wear appropriate eye protection at all times while  

handling chemicals. Depending on the types of chemicals 
you are handling, wear safety glasses with side shields, or 
goggles, or a full-face shield.

•  Wear suitable protective clothing, including gloves if  

necessary, resistant to the chemicals you are handling.

•  When preparing chemical solutions, always work in a fume 

hood that is suitable for the chemicals you are using.

•  Perform sample preparation away from the instrument  

to minimize corrosion and contamination.

•  Clean up spills immediately using the appropriate equipment 

and supplies, such as spill-cleanup kits.

Location and Space Requirements 

Space Requirements

The system should be located near the required electrical 
and gas supplies as well as the coolant supply. The roughing 
pump can be located up to a distance of 2 meters (6.5 ft.) 
from the instrument – up to 3 meters (10 ft.) using optional 
kit. There can be no more than 2 to 3 bends or couplings in 
the vacuum hose over its entire length. The diameter of the 
hose must remain at least 25 mm (1 in.) ID.  

The NexION 300 ICP-MS is designed to operate on a bench 
66-91 cm high (26-36 in. high). PerkinElmer offers a bench 
designed for the NexION 300 ICP-MS (Part No. N8141230). 
This bench is 76 cm wide x 152 cm (198 cm with shelf 
extended) long x 74 cm high (30 in. wide x 60 in. long  
[78 in. with shelf extended] x 29 in. high). This bench has 
storage bins and an acoustic barrier to muffle the sound of 
the roughing pump.

Allow space on the right and left sides of the instrument for 
the workstation or any accessories. The main air intake is 
on the left-hand side of the instrument and a minimum of 
45 cm (18 in.) clearance is required. In operation, NexION 
can be operated with the back up against a wall. Access for 
most service procedures is through the front of the instrument. 
However, some infrequent service procedures may require a 
space of at least 30 cm (12 in.) behind the instrument.

Storage Conditions

•  Ambient temperature: –20 ˚C to +60 ˚C (–4 ˚F to +140 ˚F).

•  Relative humidity 20% to 80%, without condensation.

•  Altitude: in the range 0 m to 12,000 m (sea level to  

39,370 ft.).

NOTE: When you remove the instrument from storage and before you put it 

into operation, allow it to stand for at least a day under the required 

environmental conditions.

General Laboratory Requirements

Laboratory Hygiene 

•  Keep the work area scrupulously clean to avoid con-

taminating your samples and to maintain a safe working 
environment. Clean up spilled chemicals immediately and 
dispose of them properly.

•  Do not allow waste to accumulate in the work area. 

Dispose of waste correctly.

•  Do not allow smoking in the work area. Smoking is a 

source of significant contamination and also a potential 
route for ingesting harmful chemicals.

•  Do not store, handle, or consume food in the work area.

•  Ensure that the area around, under and behind the  

instrument is clear of any dirt and dust to prevent their 
entry into the instrument’s interior, which could cause a 
negative effect on performance.

Working with Chemicals 

Some chemicals used with the instrument may be hazardous 
or may become hazardous after completion of an analysis.

•  Use, store, and dispose of chemicals in accordance with 

the manufacturer’s recommendations and the applicable 
national, state, and/or local regulations.

•  Do NOT put open containers of solvent near the instrument.

•  Store solvents in an approved cabinet (with the appropriate 

ventilation) away from the instrument.

EXPLOSIVE ATMOSPHERE. This instrument is not 
designed for operation in an explosive atmosphere.

!

WARNING

Summary of Contents for NexION 300 ICP-MS

Page 1: ...ft above sea level In addition the instrument should be located in an area that is Indoors Free of smoke dust and corrosive fumes Not prone to excessive vibration Out of direct sunlight Away from heat...

Page 2: ...quired In operation NexION can be operated with the back up against a wall Access for most service procedures is through the front of the instrument However some infrequent service procedures may requ...

Page 3: ...cart or table to allow for easy service access The system computer may be placed on the instrument bench or a separate computer table System Layout The ICP MS system consists of the main instrument ro...

Page 4: ...plosive accumulations Move or store gas cylinders only in a vertical position with the valve cap in place Locate gas cylinders away from heat or ignition sources including heat lamps Cylinders have a...

Page 5: ...l for UCT instruments only Cooling Liquid 344 14 kPa 60 2 psig 3 8 L min 1 0 gpm minimum 4 7 L min 1 25 gpm typical 5 Table 4 NexION 300 ICP MS Power Specifications Power Consumption Maximum Volt Ampe...

Page 6: ...5 Pa static pressure when the plasma is off during operation the static pressure drops to zero Both of the exhaust ports should be connected directly to flexible exhaust hoses Use the vent adapter to...

Page 7: ...umber of elbows or bends used If an excessively long duct system or a system with many bends is used a stronger blower may be necessary to provide sufficient exhaust volume at the instrument Smooth st...

Page 8: ...t common cell gases used with the NexION 300 ICP MS are ultra pure helium and anhydrous ammonia For best efficiency make sure the length of the duct that enters into the blower is a straight length at...

Page 9: ...cell gases at the working pressures listed in Table 8 The NexION 300 ICP MS is shipped with the regulator to cylinder fittings shown in Table 8 The cell gas cylinders should use this type of fitting w...

Reviews: