background image

Electrical Requirements

Power to the NexION 300 ICP-MS is to be delivered from 
one 30A/32A single-phase 200-240V dedicated electrical 
branch circuit, according to the power specifications in Table 4.  
Table 5 provides the electrical supply requirements and 
approximate power consumption of the peripherals. If the 
power line is unstable, fluctuates or is subject to surges, 
additional control of the incoming power may be required. 

The vacuum roughing pump is provided with a mains supply  
plug suitable for the country of installation (shown in Figure 4)  
and must be connected to a separate branch circuit/wall 
outlet. It requires one 15A single-phase 200-240V outlet 
– see Table 5. See Figure 3 (Page 4) for the location and 
lengths of hoses, lines, cords and cables.   

Facilities Requirements

Table 3 provides information on the gas and liquid services required for the NexION 300 ICP-MS. Tables 4 and 5 show the 
electrical supply requirements and approximate power consumption of the NexION 300 ICP-MS and its major accessories.

Table 3.  Gas and Liquid Services Required for the NexION 300 ICP-MS. 
Gases 

Operating pressure 

Flow at operating pressure

Argon  

586 - 690 kPa (85 – 100 psig) min-max 

14-20 L/min (typical) 

Purity see page 8 

 

     

Ammonia ≥ 99.999% pure 

48 ±14 kPa (7 ±2 psig) operating 

0.6 mL/min (typical) 

(for UCT instruments only) 

 

    

Helium ≥ 99.9999% pure 

290 ±14 kPa (42 ±2 psig) operating 

5 mL/min (typical) 

(for UCT instruments only) 

 

    

 

Cooling Liquid 

344 ±14 kPa (60 ±2 psig) 

3.8 L/min (1.0 gpm) minimum 

 

 

4.7 L/min (1.25 gpm) typical

5

Table 4.  NexION 300 ICP-MS Power Specifications.  
Power Consumption: 

 

Maximum Volt Amperes (total) 

3720VA

Maximum Continuous Current  

22A

Voltage Amplitude Specification:   
Operating Voltage 

200 - 240

Allowable Voltage Variance 

±10%

Maximum Allowable Percent Sag 

5%

Maximum Allowable Percent Swell  5%
Phase 

Single phase only

Frequency Specification: 

 

Operating Frequency 

50/60 Hz

Allowable Frequency Variance 

±1 Hz

Waveform Specification: 

 

Maximum Supply Voltage  

Total Distortion 

5%

Maximum Supply Voltage  

Distortion by Single Harmonic 

3%

Figure 4.  Vacuum roughing pump mains supply plugs.

North America 
Japan 
NEMA 6-15P 
N8145006

Europe 
CEE 7 
“Schuko” 
N8145007

Switzerland 
N8145009

United Kingdom 
BS 1363 
N8145008

Rest of World 
No plug 
N8145010

MAGNETIC SUSCEPTIBILITY. Do NOT place NexION 
300 ICP-MS close to any other instrumentation 
or equipment that emits high magnetic fields. 
External magnetic field strength must not exceed 
10 Gauss at NexION 300 ICP-MS.

•  Use only approved regulators and hose connectors. Left-

hand thread fittings are used for fuel gas tank connections 
whereas right-hand fittings are used for oxidant and  
support gas connections.

•  Arrange gas hoses where they will not be damaged or 

stepped on and where things will not be dropped on 
them.

•  Perform periodic gas leak tests by applying a soap solution 

to all joints and seals.

•  It is strongly recommended that Universal Cell Tech-

nology

 (UCT) gases be installed in a gas cabinet with 

adequate ventilation.

!

WARNING

Summary of Contents for NexION 300 ICP-MS

Page 1: ...ft above sea level In addition the instrument should be located in an area that is Indoors Free of smoke dust and corrosive fumes Not prone to excessive vibration Out of direct sunlight Away from heat...

Page 2: ...quired In operation NexION can be operated with the back up against a wall Access for most service procedures is through the front of the instrument However some infrequent service procedures may requ...

Page 3: ...cart or table to allow for easy service access The system computer may be placed on the instrument bench or a separate computer table System Layout The ICP MS system consists of the main instrument ro...

Page 4: ...plosive accumulations Move or store gas cylinders only in a vertical position with the valve cap in place Locate gas cylinders away from heat or ignition sources including heat lamps Cylinders have a...

Page 5: ...l for UCT instruments only Cooling Liquid 344 14 kPa 60 2 psig 3 8 L min 1 0 gpm minimum 4 7 L min 1 25 gpm typical 5 Table 4 NexION 300 ICP MS Power Specifications Power Consumption Maximum Volt Ampe...

Page 6: ...5 Pa static pressure when the plasma is off during operation the static pressure drops to zero Both of the exhaust ports should be connected directly to flexible exhaust hoses Use the vent adapter to...

Page 7: ...umber of elbows or bends used If an excessively long duct system or a system with many bends is used a stronger blower may be necessary to provide sufficient exhaust volume at the instrument Smooth st...

Page 8: ...t common cell gases used with the NexION 300 ICP MS are ultra pure helium and anhydrous ammonia For best efficiency make sure the length of the duct that enters into the blower is a straight length at...

Page 9: ...cell gases at the working pressures listed in Table 8 The NexION 300 ICP MS is shipped with the regulator to cylinder fittings shown in Table 8 The cell gas cylinders should use this type of fitting w...

Reviews: