background image

Exhaust and Ventilation Requirements

 

The NexION 300 ICP-MS has two separate exhaust ports. 

Both of the NexION 300 ICP-MS exhaust ports are located 
on the top of the instrument (see Figure 6). The torch box 
exhaust port is located 46.5 cm (18.5 in.) from the right side 
of the instrument. The RF generator exhaust port is located 
67.5 cm (27 in.) from the left side of the instrument.

The torch box exhaust port exhausts the following:

•  Plasma heat and fumes

•  Vacuum pump exhaust – including cell gases on Universal 

Cell Technology

 (UCT) instruments

•  Cell gas assembly manual vent/purge switch

The torch box exhaust venting system is required to remove 
combustion fumes and vapors from the torch housing, and 
to remove reaction cell gas. Exhaust venting is important for 
four reasons:

•  It protects laboratory personnel from toxic vapors that 

may be produced by some samples.

•  It minimizes the effects of room drafts and the laboratory 

atmosphere on ICP torch stability.

•  It helps protect the instrument from corrosive vapors 

which may originate from the samples.

•  It removes dissipated heat which is produced by the ICP torch.

The torch exhaust port always has 0.5 inches of water (125 Pa) 
static pressure. The RF generator exhaust port has 0.5 inches 
of water (125 Pa) static pressure when the plasma is off – 
during operation, the static pressure drops to zero. Both of 
the exhaust ports should be connected directly to flexible 
exhaust hoses. Use the vent adapter to attach the roughing 
pump exhaust hose to the torch box exhaust port.  

Table 5.  Electrical Requirements for NexION 300 ICP-MS 

Peripherals.
Equipment Voltage(AC)  Power
Computer 

100-127V/200-240V,  

800W  

 

50/60Hz 

typical

Printer 

100-127V/220-240V,  

800W  

 

50/60Hz 

typical

Roughing Pump 

200-240V, 50/60Hz 12A  1500W

Cooling System: 
Heat Exchanger 

208-240V, 50/60Hz 

660W  

(PolyScience

®

 3370)   120V, 60Hz 5.5A

or
Refrigerated chiller  

208-230V, 60Hz 12.2A 

2500W 

(PolyScience

®

 6106PE)  240V, 50Hz 12.2A

Mains Connection

The instrument is shipped with one 2.4 m (8 ft.) mains cord 
terminated by an IEC 60309 connector rated 30A by UL (North 
America) and 32A by VDE (International) for 250V as shown in 
Figure 5.

6

Figure 5.  IEC 60309 connector.

Figure 6.  Location of exhaust ports.

EXPLOSIVE ATMOSPHERE. The use of ICP-MS 
instruments without adequate ventilation to  
outside air may constitute a health hazard.  
For example, the combustion of halogenated 
hydrocarbons produces toxic vapors. When  
ammonia is used as a Dynamic Reaction Cell  
gas, it can be present at very trace amounts  
in the exhaust during purging of the gas lines. 
Extreme care should be taken that exhaust  
gases are vented properly.

!

WARNING

67.5 cm

(27 in.)

46.5 cm

(18.5 in.)

RF Generator

Exhaust Port

Torch Box

Exhaust Port

Summary of Contents for NexION 300 ICP-MS

Page 1: ...ft above sea level In addition the instrument should be located in an area that is Indoors Free of smoke dust and corrosive fumes Not prone to excessive vibration Out of direct sunlight Away from heat...

Page 2: ...quired In operation NexION can be operated with the back up against a wall Access for most service procedures is through the front of the instrument However some infrequent service procedures may requ...

Page 3: ...cart or table to allow for easy service access The system computer may be placed on the instrument bench or a separate computer table System Layout The ICP MS system consists of the main instrument ro...

Page 4: ...plosive accumulations Move or store gas cylinders only in a vertical position with the valve cap in place Locate gas cylinders away from heat or ignition sources including heat lamps Cylinders have a...

Page 5: ...l for UCT instruments only Cooling Liquid 344 14 kPa 60 2 psig 3 8 L min 1 0 gpm minimum 4 7 L min 1 25 gpm typical 5 Table 4 NexION 300 ICP MS Power Specifications Power Consumption Maximum Volt Ampe...

Page 6: ...5 Pa static pressure when the plasma is off during operation the static pressure drops to zero Both of the exhaust ports should be connected directly to flexible exhaust hoses Use the vent adapter to...

Page 7: ...umber of elbows or bends used If an excessively long duct system or a system with many bends is used a stronger blower may be necessary to provide sufficient exhaust volume at the instrument Smooth st...

Page 8: ...t common cell gases used with the NexION 300 ICP MS are ultra pure helium and anhydrous ammonia For best efficiency make sure the length of the duct that enters into the blower is a straight length at...

Page 9: ...cell gases at the working pressures listed in Table 8 The NexION 300 ICP MS is shipped with the regulator to cylinder fittings shown in Table 8 The cell gas cylinders should use this type of fitting w...

Reviews: