background image

Introduction

PerkinElmer

®

 ICP-MS instruments are complete systems 

with the exception of the following items which must 
be provided by the customer: electrical power, exhaust 
vents, argon gas supplies with approved regulator, cell 
gas supply, and coolant system. 

Required Environmental Conditions

The laboratory environment in which the NexION

®

 300 

ICP-MS instrument is installed should meet the following 
conditions:

•  The room temperature should be between 15 and 

30 ˚C (59-86 ˚F) with a maximum rate of change of 
3 ˚C (5 ˚F) per hour.

•  The relative humidity should be between 20 and 

80%, non-condensing. For optimum performance, 
the room temperature should be controlled at  
20 ±2 ˚C (68 ±3.6 ˚F), and the relative humidity 
should be between 35 and 50%.

•  The instrument is certified for operation at elevations up 

to 2000 meters (6562 ft.) above sea level.

In addition, the instrument should be located in an 
area that is:

- Indoors

- Free of smoke, dust and corrosive fumes

- Not prone to excessive vibration

- Out of direct sunlight

- Away from heat radiators

In order to minimize contamination problems, a 
dust-free environment is necessary. For ultra-trace 
techniques, environmental contamination becomes a 
limiting factor in the analysis. To quantitate ubiquitous 
elements such as Fe, Ca, K, Na, etc. below 1 ppb (µg/L),  
a class 1000 environment is necessary for sample  
preparation and analysis. This is not an indication of  
the performance limitations of the instrument, but  
a recommendation for an ultra-clean environment.

The NexION 300 ICP-MS can be installed into a mobile 
laboratory if vibration is isolated. 

P R E P A R I N G   Y O U R   L A B

ICP-Mass Spectrometry

Preparation Checklist

•  Environmental conditions
•  Electrical requirements
•  Space requirements
•  Exhaust ventilation
•  Coolant requirements
•  Argon gas requirements
•  Cell gas requirements

NexION 300 ICP-MS

Summary of Contents for NexION 300 ICP-MS

Page 1: ...ft above sea level In addition the instrument should be located in an area that is Indoors Free of smoke dust and corrosive fumes Not prone to excessive vibration Out of direct sunlight Away from heat...

Page 2: ...quired In operation NexION can be operated with the back up against a wall Access for most service procedures is through the front of the instrument However some infrequent service procedures may requ...

Page 3: ...cart or table to allow for easy service access The system computer may be placed on the instrument bench or a separate computer table System Layout The ICP MS system consists of the main instrument ro...

Page 4: ...plosive accumulations Move or store gas cylinders only in a vertical position with the valve cap in place Locate gas cylinders away from heat or ignition sources including heat lamps Cylinders have a...

Page 5: ...l for UCT instruments only Cooling Liquid 344 14 kPa 60 2 psig 3 8 L min 1 0 gpm minimum 4 7 L min 1 25 gpm typical 5 Table 4 NexION 300 ICP MS Power Specifications Power Consumption Maximum Volt Ampe...

Page 6: ...5 Pa static pressure when the plasma is off during operation the static pressure drops to zero Both of the exhaust ports should be connected directly to flexible exhaust hoses Use the vent adapter to...

Page 7: ...umber of elbows or bends used If an excessively long duct system or a system with many bends is used a stronger blower may be necessary to provide sufficient exhaust volume at the instrument Smooth st...

Page 8: ...t common cell gases used with the NexION 300 ICP MS are ultra pure helium and anhydrous ammonia For best efficiency make sure the length of the duct that enters into the blower is a straight length at...

Page 9: ...cell gases at the working pressures listed in Table 8 The NexION 300 ICP MS is shipped with the regulator to cylinder fittings shown in Table 8 The cell gas cylinders should use this type of fitting w...

Reviews: