25
O
2
asis User Guide
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m
Power frequency magnetic fields from
common appliances in the home are not
expected to affect the device.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital envi-
ronment. Keep the O
2
asis Humidifier away
from sources of high levels of power line
magnetic fields (in excess of 30 A/m) to
reduce the likelihood of interference.”
NOTE: UT is the A/C. mains voltage prior to application of the test level.
Electromagnetic Immunity
The O
2
asis Humidifier is intended for use in the electromagnetic environment specified be-
low. The user of the O
2
asis Humidifier should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment – Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
6 V rms in ISM
and amateur radio
bands
10 V/m
The O2asis Humidifier
is suitable for the
electromagnetic environment
of typical homes, commercial
or hospital settings.
DISPOSAL
•
Base Unit:
This device should be disposed of separately from standard mu-
nicipal waste. Dispose of in accordance with local regulations for recycled
electronic goods.
•
Water Well Consumable Cartridge:
Dispose of the Water Well consumable
cartridge in accordance with local plastic recycling guidelines.
TROUBLESHOOTING
Below is a list of possible problems you may have with the device and the ac-
tions to take if these problems occur.
Problem
Possible Cause
Action to Take
The three LEDs
on the base unit
are flashing.
There is an error
with the device.
Unplug the device from the wall outlet and
then plug back in. Check to see if the water
well is empty. If empty, refill with distilled water.
Also check to make sure that the consumable
cartridge is properly secured to the base and
that the hoses and power connector are all
properly connected to their respective ports. If
you have verified all of the above are correct,
unplug the device and then please contact the
care provider, licensed distributor, or O2asis to
return the device.
PP-19-1 REV. 15