O

2

asis User Guide

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

30A/m

Power frequency magnetic fields from 
common appliances in the home are not 
expected to affect the device.
Power frequency magnetic fields should be 
at levels characteristic of a typical location 
in a typical commercial or hospital envi-
ronment. Keep the O

2

asis Humidifier away 

from sources of high levels of power line 
magnetic fields (in excess of 30 A/m) to 
reduce the likelihood of interference.”

NOTE: UT is the A/C. mains voltage prior to application of the test level.

Electromagnetic Immunity

The O

2

asis Humidifier is intended for use in the electromagnetic environment specified be-

low. The user of the O

2

asis Humidifier should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test 
Level

Compliance Level

Electromagnetic
Environment – Guidance

Conducted RF 
IEC 61000-4-6

Radiated RF 
IEC 61000-4-3

3 Vrms
150 kHz to
80 MHz

3 V/m
80 MHz to
2.5 GHz

3 Vrms
6 V rms in ISM 
and amateur radio 
bands

10 V/m

The O2asis Humidifier 
is suitable for the 
electromagnetic environment 
of typical homes, commercial 
or hospital settings.

DISPOSAL

• 

Base Unit: 

This device should be disposed of separately from standard mu-

nicipal waste. Dispose of in accordance with local regulations for recycled 
electronic goods.

• 

Water Well Consumable Cartridge: 

Dispose of the Water Well consumable 

cartridge in accordance with local plastic recycling guidelines.

TROUBLESHOOTING

Below is a list of possible problems you may have with the device and the ac-
tions to take if these problems occur.

Problem

Possible Cause

Action to Take

The three LEDs 
on the base unit 
are flashing.

There is an error 
with the device.

Unplug the device from the wall outlet and 
then plug back in. Check to see if the water 
well is empty. If empty, refill with distilled water. 
Also check to make sure that the consumable 
cartridge is properly secured to the base and 
that the hoses and power connector are all 
properly connected to their respective ports. If 
you have verified all of the above are correct, 
unplug the device and then please contact the 
care provider, licensed distributor, or O2asis to 
return the device.

PP-19-1 REV. 15

Summary of Contents for O2asis

Page 1: ...Personal Humidifier USER GUIDE PP 19 1 REV 15...

Page 2: ...und to be defective within the stated warranty period Should the O2asis base unit or the replacement water well fail prematurely the replacement base unit is warranted for the remainder of the origina...

Page 3: ...djusting the Heat Level 13 Pediatric Patients 13 Use Without Electrical Power 14 Replacing the Water Well Consumable Cartridge 15 Checking for an empty Water Well 17 Cleaning and Disinfection 18 Regul...

Page 4: ...the power cord in a way that will avoid anyone tripping or anything otherwise interfering with it Do not position the device in a way that would make it difficult to access the power supply connection...

Page 5: ...er Do not attempt to use with or connect to other medical equipment This device is not intended for connection to a medical pipeline This device is not intended for use with patients whose upper airwa...

Page 6: ...ryloni trile which is known in the state of California to cause cancer For more information go to www p65warnings ca gov INCLUDED WITH THE SYSTEM The O2asis Humidifier comes with the following parts N...

Page 7: ...wer Cord Connector The connector that accepts the male connector at the end of the power cord 6 Heated Line The heated tube attached to the Water Well cartridge that heats the oxygen to prevent water...

Page 8: ...ns Heated Line Power Levels Water Heater Temperature Levels Water Heater power turned off when Heated Line is not detected LED Indicator Lights 3 LED Indicator Lights low medium high User Interfaces T...

Page 9: ...e For additional reservoir durations please refer to the Performance section of this manual Note The device will detect when the Water Well is empty and will shut down The three LED lights will flash...

Page 10: ...o prevent over filling the pump should not be used while the Water Well is disconnected from the Base Unit Confirm that the electrical port is dry and reattach the device to an electrical outlet If th...

Page 11: ...ment Note Make sure that the Water Well consumable cartridge is securely latched to the base unit before plugging in the power cord Note The O2asis Humidifier device will continue to humidify the airf...

Page 12: ...steps Warning Use only the cannula included with the device Use of other cannulas can degrade the performance of the device 1 Remove the cannula from its plastic bag 2 Connect the tube end to the conn...

Page 13: ...ed oxygen flow as shown in the table below 2 Above 2lpm of oxygen flow begin your oxygen therapy on the first level of heat When the device receives power the left most LED light will illuminate signi...

Page 14: ...in Adjusting the Heat Level sec tion of this document Condensation Regularly check for visual condensation in the cannula If excessive amounts of condensation are present reduce the heat setting to t...

Page 15: ...age device To replace the Water Well consumable cartridge complete the following steps 1 Unplug the power adapter cord from the latch side of the device 2 Unplug the oxygen source line from the oxygen...

Page 16: ...ging 8 Insert the non port side of the cartridge into the button side of the base and push in until the cartridge sits flush with the base 9 Swing the base latch up and snap it onto the cartridge to l...

Page 17: ...d can become contaminated with bodily fluids If visible contamination is present or performance degradation is suspected dispose of and replace the Water Well consumable cartridge Warning Do not touch...

Page 18: ...oap Wipe all exterior surfaces If visible soil is present use a wetted soft nylon bris tle brush to remove visible soil paying close attention to seams and crevices in the device Wipe again with dish...

Page 19: ...ganic solvents or abrasive cleaners These types of cleaners may create toxic gasses when heated and may damage the device Warning Water fill pump is intended for single patient use only The Water Fill...

Page 20: ...hysical Dimensions 187 x 102 x 30mm 7 4 x 4 0 x 1 2 in Not including 1118mm 44 in long heated line Weight 0 7 lbs 65lbs 295 g empty 10 95 lbs 430 g full Distilled water reservoir approx capacity 135 m...

Page 21: ...ion This device is an US FDA cleared Class II medical device This device is considered an IEC Class II continuous operation ordinary equipment device The heated line with cannula is considered an IEC...

Page 22: ...w Rate lpm liters per minute Leakages ml milliliters Oxygen Supply Flow Rate Range 0 5 lpm to 6 lpm Maximum Oxygen supply flow rate 6 lpm ESTIMATED HUMIDIFICATION OUTPUT Dry Oxygen Flow Rate Heat Sett...

Page 23: ...Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the O2asis Humidifier Syst...

Page 24: ...domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Immunity Electromagnetic Immunity The O2asis Humidifier is intended for use in...

Page 25: ...io bands 10 V m The O2asis Humidifier is suitable for the electromagnetic environment of typical homes commercial or hospital settings DISPOSAL Base Unit This device should be disposed of separately f...

Page 26: ...snug If the device still does not provide the proper air flow please contact the care provider licensed distributor or O2asis to return the device Problem Possible Cause Action to Take The temperatur...

Page 27: ...ng comfort anymore please purchase a new Water Well to restore full comfort Problem Possible Cause Action to Take The device does not seem to be operating correctly You accidentally spilled water onto...

Page 28: ...1001 New Hampshire Ave Lakewood NJ 08701 USA Toll Free 1 800 337 3762 www o2asishumidifier com...

Reviews: