19
O
2
asis User Guide
Reprocessing and Disinfection Between Patients
The O2asis Humidifier Base must always be cleaned and disinfected between
patients. The entire Water Well consumable cartridge, including its blue
heated
line
, is disposable and no disinfection is required. Follow these steps to ensure
a clean and disinfected device:
1
Unplug the O
2
asis Humidifier Base from the power supply.
2
Remove and discard the Cannula ( single-patient use).
3
Remove and discard the Water Well consumable cartridge ( single-
patient use).
4
Examine for visible soil. If visible soil is present, use a wetted soft nylon
bristle brush or Super Sani-Cloth® to remove all visible soil, paying close
attention to seams and crevices in the device.
5
Wet all surfaces of the base unit with another Super Sani-Cloth®. Keep
the surface wet for at least two minutes. Use additional Super Sani-
Cloth® if needed.
6
Allow the device to air dry completely.
7
If Power Supply is exposed to the patient, disconnect from the O2asis
device and wall. Repeat steps 4 through 6, and let thoroughly dry.
Warning:
Do not
use organic solvents or abrasive cleaners. These types of
cleaners may create toxic gasses when heated and may damage the device.
Warning:
Water fill pump is intended for single patient use only. The
Water Fill Pump should not be used between multiple patients. It cannot be
reprocessed. If the pump may have been contaminated, replace or provide
patient with new pump.
The Super Sani-Cloth® Germicidal Disposable Wipe is a premoistened nonwo-
ven durable wipe containing a quaternary ammonium chloride/alcohol-based
solution. It is recommended for use in hospitals and other critical care areas
where the control of the hazards of cross-contamination between treated
surfaces is required, and it is designed to be compatible with hard nonporous
surfaces and equipment made of plastic, Formica® laminate, glass and more.
Some organisms are removed from the surface by thoroughly wiping the sur-
face with the wipe. Most remaining organisms are killed within two (2) minutes
by exposure to the liquid in the wipe.
The Super Sani-Cloth Wipe is registered with the Environmental Protection
Agency (EPA) under (EPA #9480-4). In accordance with 21 CFR 880.6890,
the Super Sani-Cloth Wipe is regulated by the FDA as a Class I, 510(k) Exempt,
General Purpose Disinfectant under Product Code (LRJ).
PP-19-1 REV. 15