50 . STA 8000 Installation and Hardware Guide
Once Pyris Software has been configured for the STA 8000, the STA 8000 Application can be
started from Pyris Manager.
Starting the Pyris Manager
1.
Select
Pyris Manager
from the Pyris Software group in Programs on the Start menu.
The Pyris Manager is displayed on the screen.
2.
Select
STA 8000
from the Pyris Manager.
The STA 8000 control panel is displayed.
3.
Heat the sensor to approximately 150 °C to burn off any excess moisture.
Enter 150 in the box beneath the Go To Temp button
on the control panel
and click the button.
4.
After 10–15 minutes, cool the analyzer to ambient temperature.
The configuration for the STA 8000 is complete. The analyzer is ready to be calibrated.