Install an STA 8000 . 47
When using tap water, the temperature should not change more than 0.4 °C/hour.
CAUTION
ATTENTION
The cooling jacket of the STA 8000 will NOT accept pressure. Use the
lowest flow rate that provides sufficient cooling for your experiment to
minimize the pressure in the system.
La gaine de refroidissement du STA 8000 n'acceptera PAS la pression.
Utilisez le débit le plus bas qui procure un refroidissement suffisant pour
votre expérience afin de minimiser la pression dans le système.
NOTE: It is essential to maintain a very constant flow rate of water. The liquid should be
free of air bubbles. Changes in the liquid flow rate or the presence of air bubbles will
reduce the quality of the measurement signal.