3.2.3
Replacing a Blown Fuse
If one of the two fuses of the Operetta has blown, it can be exchanged by the user.
Danger!
Electric Shock!
l
Switch off the Operetta and disconnect the power cable before
changing the fuse(s).
Caution!
A defective fuse may only be replaced by exactly the same part from
the following manufacturer:
LITTLEFUSE 218000 Series
5 × 20 mm Time Lag Fuse (Slo-Blo)
15 A / 250 V (US), 8 A / 250 V (EU).
1. Switch off the Operetta and disconnect the power cable.
2. Pull out the fuse holder.
3. Replace the blown fuse(s).
4. Close the fuse holder and push it into its former position.
5. Reconnect the power cable.