Send Back (Fax or Email)
http://www.perkinelmer.com/AboutUs/ContactUs/ContactUs
Date:
To:
PerkinElmer
Fax Number:
Customer:
Subject:
Installation of Opera Phenix Plus
I have reviewed the Site Readiness Instructions and confirm that:
Meets Requirements
o
The available space at the installation location meets the space requirements
of the Opera Phenix Plus (including upgrade components, if applicable). See
chapter 2.3 .
o
Additional space for unpacking/installing the Opera Phenix Plus is available as
specified at the installation location (only during delivery). See chapter 2.3 .
o
The laboratory table has a carrying capacity of at least 400 kg (882 lb).
o
The laboratory meets the temperature (20 °C to 25 °C) and humidity (less than
70 % non condensing) requirements.
o
The room is well ventilated and free from dust.
o
At least three electrical outlets are available within 1.5 m (59 in) of the
installation location.
o
Electrical supply has a frequency of 50 or 60 Hz and a voltage in the range of
100 to 120 V or 220 to 240 V.
o
*
* Only with optional ECO upgrade:
CO
2
supply is provided by the customer as
specified in the gas requirements. See chapter 2.7 .
o
The internal transportation of the shipment containers from the customer's
receiving department right to its desired room position is guaranteed.
o
Any customer-supplied PC for Harmony office installation meets the
requirements described. See chapter 3.3 .
o
The packing material will be disposed of by the customer.
o
The local IT department has agreed to the integration of the Opera Phenix Plus
system (including the Harmony PC running with Windows® 10).
The preferred solution for the network integration will be:
o
No integration
o
Simple Integration (chapter 3.1.2 )
o
Office PC – big (chapter 3.1.3.2 )
o
Office PC – small (chapter 3.1.3.1 )
o
Router (chapter 3.1.3.3 )
o
The required hardware for the network integration will meet the requirements
and will be available for the installation. See chapter 3.1 .
Sincerely,
Name
Telephone Number
4 Confirmation Fax/Email
44
Opera Phenix Plus – Site Readiness
Summary of Contents for OPERA PHENIX PLUS
Page 1: ...OPERA PHENIX PLUS Site Readiness...
Page 4: ......
Page 5: ...Opera Phenix Plus Site Readiness 5...
Page 6: ...1 Introduction 1 Introduction 6 Opera Phenix Plus Site Readiness...
Page 8: ......
Page 9: ...2 Requirements 2 Requirements Opera Phenix Plus Site Readiness 9...
Page 27: ...3 IT Policy 3 IT Policy Opera Phenix Plus Site Readiness 27...
Page 43: ...4 Confirmation Fax Email 4 Confirmation Fax Email Opera Phenix Plus Site Readiness 43...
Page 45: ......