Country Kit
Used for
Keyboard
layout
Power plug
Plug type/norm
HH14001610
(JP)
Japan
US QWERTY
JIS C 8303
498 GJ
HH14001611
(Export)
European
Export
UK QWERTY
KC-015A
IEC 60884-1
HH14001612
(CH)
Switzerland,
Liechtenstein
DE QWERTZ,
AZERTY
SEV 1011
SEV 6534/2
HH14001613
(Nordics)
Sweden,
Finland,
Denmark
US QWERTY
KC-015A
IEC 60884-1
HH14001614
(East Europe)
Poland,
Hungary
US QWERTY
KC-015A
IEC 60884-1
2.5
Environmental Requirements
l
Operating conditions:
o
+20 °C to +25 °C (68 °F to 77 °F)
o
Relative humidity < 70 %, non-condensing
l
Storage temperature: 10 °C to 40 °C (50 °F to 104 °F)
l
Indoor use only
l
A climate-controlled, clean and dust-free laboratory is recommended
l
Atmospheric pressure up to 2000 m altitude
l
Vibration-free work surface
2.6
Biosafety
Danger! Biohazard!
This device is designed for analyzing biological substances at
biosafety level 1 and 2 (BSL-1 or BSL-2).
l
Do not use more hazardous substances due to limited
possibilities of decontamination.
l
Observe the safety instructions for the substances used.
l
Observe the instructions for decontamination and cleaning.
l
Always make sure to comply with national and international
laboratory safety instructions.
l
Always wear your personal safety equipment.
2 Requirements
Opera Phenix Plus – Site Readiness
21
Summary of Contents for OPERA PHENIX PLUS
Page 1: ...OPERA PHENIX PLUS Site Readiness...
Page 4: ......
Page 5: ...Opera Phenix Plus Site Readiness 5...
Page 6: ...1 Introduction 1 Introduction 6 Opera Phenix Plus Site Readiness...
Page 8: ......
Page 9: ...2 Requirements 2 Requirements Opera Phenix Plus Site Readiness 9...
Page 27: ...3 IT Policy 3 IT Policy Opera Phenix Plus Site Readiness 27...
Page 43: ...4 Confirmation Fax Email 4 Confirmation Fax Email Opera Phenix Plus Site Readiness 43...
Page 45: ......