Routine Maintenance and Troubleshooting . 89
Figure 34 Dismantling or assembling the flowcell
3.
When the flowcell is dismantled, use a lens tissue and a suitable solvent, for example
heptane, to clean the flowcell.
WARNING
Do not use acid to clean the windows, as toxic fumes will be given off.
Take care not to crease the wedged PTFE spacer during cleaning as this
may reduce its effectiveness at suppressing interference fringes, and may
also alter the pathlength.
4.
Refit the parts of the cell in the following order:
–
The neoprene gasket
–
The back window
–
The former
–
The PTFE spacer (with the thin edge of the wedge towards the top of the cell)
–
The front window
–
The PTFE gasket
5.
Align the two holes in the PTFE gasket with the two holes at the ends of the channels in
the front window.