42 . OilExpress System User's Guide
Positioning the OilExpress System
WARNING
The Autosampler is a heavy instrument with a basic weight of
approximately 30 kg.
Two people are required for safe handling.
Consult the local codes of practice issued by safety advisors before
attempting to lift it.
Take care not to injure yourself or others, or to drop the instruments.
1.
Position the Liquid Autosampler on the bench as shown below (Figure 15).
Leave a gap of 0.5–1 cm between the Autosampler and the spectrometer. Ensure that
you have space for the PC either to the right side of the spectrometer, or to the left side
of the Autosampler.
Figure 15 The OilExpress System positioned for use
2.
Align the Autosampler and the spectrometer so that the rear of the injection wash port
and the sample shuttle are parallel.
The spectrometer protrudes approximately 5 cm in front of the Autosampler when
correctly aligned.
Make sure that there is access to power and a waste container.
3.
Place the PC in a suitable location.
4.
Remove the Velcro straps and plastic cable ties around the Z-rack.
5.
Remove any additional packaging and support from the Autosampler and make sure
that the arm is able to move freely on all axes.
Summary of Contents for OilExpress System
Page 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 23: ...Warnings and Safety Information ...
Page 36: ...36 OilExpress System User s Guide ...
Page 37: ...Unpacking and Installation ...
Page 69: ...Background to Used Oil Analysis ...
Page 75: ...Routine Maintenance and Troubleshooting ...
Page 96: ...96 OilExpress System User s Guide ...
Page 97: ...Appendices ...