50 . OilExpress System User's Guide
Installing the Syringe
The syringes cannot be installed until the OilExpress software has been installed.
1.
Switch on the Autosampler.
2.
Start the AutoSampler Utilities.
Select
Start
,
All Programs
,
PerkinElmer Applications
,
OilExpress
and then click
AutoSampler Utilities
.
3.
Log in as an Administrator.
The User Name is
Administrator
and the Password is
Administrator
.
4.
Select
Pump Diagnostics
from the File menu.
5.
Click
Change Syringe
.
6.
Assemble the syringe (L1200552) by placing the plunger in syringe barrel and putting
the syringe clamp over the Luer lock fitting.
7.
Unscrew the pin from the plunger shaft.
8.
Mount the syringe plunger on the plunger shaft. Line up the barrel with the Luer lock
fitting. Screw the barrel onto the valve fitting while pushing upward slightly.
9.
Replace the plunger pin.
10.
Rotate the syringe clamp clockwise then tighten the clamp screw.
11.
Replace the plunger screw at the base of the syringe plunger.
CAUTION
When mounting the syringe on the valve, you must assist the Luer lock
threads by pushing upward. Otherwise they may become stripped.
Summary of Contents for OilExpress System
Page 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 23: ...Warnings and Safety Information ...
Page 36: ...36 OilExpress System User s Guide ...
Page 37: ...Unpacking and Installation ...
Page 69: ...Background to Used Oil Analysis ...
Page 75: ...Routine Maintenance and Troubleshooting ...
Page 96: ...96 OilExpress System User s Guide ...
Page 97: ...Appendices ...