20 . OilExpress System User's Guide
Analyzer Measurement
Analyzer Sample Compartment
WARNING
DO NOT use the OilExpress System when the sample compartment lid is
fitted. Flammable vapor may seep through instrument seals and gaskets
and accumulate under the lid, where it could cause a fire or explosion
hazard.
CAUTION
The sample compartment windows may be composed of a hygroscopic
material, such as KBr; which, although coated, can be damaged by high
levels of humidity. Refer to the documentation supplied with your
spectrometer for details.
If you spill liquid in the sample compartment, wipe it up quickly.
CAUTION
Relative humidity higher than 80% can damage the windows of the
sample compartment.
The sample compartment (Figure 10) is located at the front of the instrument.
Figure 10 Sample compartment of a Frontier IR System
•
The infrared beam enters the compartment through an aperture on the left. After
passing through the sample, it enters the detector area through an aperture on the
right side of the sample compartment.
•
A sample shuttle is installed in the sample compartment. The sample shuttle is used to
move the flowcell of the Liquid Autosampler in and out of the beam.
Summary of Contents for OilExpress System
Page 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 23: ...Warnings and Safety Information ...
Page 36: ...36 OilExpress System User s Guide ...
Page 37: ...Unpacking and Installation ...
Page 69: ...Background to Used Oil Analysis ...
Page 75: ...Routine Maintenance and Troubleshooting ...
Page 96: ...96 OilExpress System User s Guide ...
Page 97: ...Appendices ...