Introduction . 9
Figure 2. Energy Absorption
Once a sample absorbs light energy, the intensity of incoming light P
0
(the energy from the
light source) reduces to P by losing some of its energy;
therefore, we define the transmittance of a sample as follows:
Figure 3. Sample Absorption
As a result of absorption by a sample, transmittance is defined as;
T = P / P
0
And Absorbance is defined as:
A = -log T = log( P
0
/ P )Absorbance is related to c (concentration) and b (pathlength of
light) according to the Lambert-Beer law and the equation;
A=εbc
Where,
ε, the molar absorption coefficient of a particular molecule, varies with the
wavelength.
Although Beer's law has several limitations, it is widely applied to quantitative analysis.
Summary of Contents for Lambda 365
Page 1: ...LAMBDA 365 Users Guide UV SPECTROMETERS ...
Page 4: ...4 Lambda 365 Users Guide ...
Page 5: ...Introduction 5 Introduction ...
Page 19: ...Safety Practices 19 Safety Practices ...
Page 30: ...30 Lambda 365 Users Guide ...
Page 31: ...Lambda 365 Installation 31 Lambda 365 Installation ...
Page 35: ...UV Express Software Installation 35 UV Express Software Installation ...
Page 37: ...Interface Setup 37 3 UV Express InstallShield Wizard starts Select Next ...
Page 41: ...Interface Setup 41 Interface Setup ...
Page 43: ...Interface Setup 43 5 Double click USBCom CDM_20828 6 Select Extract ...
Page 46: ...46 Lambda 365 Users Guide ...
Page 47: ...Setting USB Serial Ports 47 Setting USB Serial Ports ...
Page 48: ...48 Lambda 365 Users Guide Procedure 1 Select Computer Properties 2 Click on Device Manager ...
Page 53: ...Operating Procedure 53 Operating Procedure ...
Page 55: ...Accessories 55 Accessories ...
Page 62: ...62 Lambda 365 Users Guide ...