FLEX 2100
21
Sample rate:
The sample rate defines the refreshment speed of the weighing signal.
For example:
If a weighing unit is filling bottles of milk with 1600 gram and the Sample rate is set to 1600
samples/sec the unit is able to fill this bottle in 1 second and see every gram of it, so a
accuracy of +/- 1 gram has become possible.
If the same unit is set to a sample rate of 800 samples/sec the unit is still able to fill the
bottle in 1 second but it now will see only the halve of the grams, so now a accuracy of +/-
2 gram has become possible.
To get the same accuracy with 800 sample/sec as with 1600 samples/sec the machine
should be set to fill the bottles within 2 seconds.
Sample rate can be set at 10, 20, 25, 50, 100, 200, 400 , 800 or 1600 samples/sec.
Note: The accuracy of a weigher is always depending on the chosen load cells and the
mechanical construction.
Certified text:
Certified text is only used when the weigher is certified and sealed by an authorized
person.
The text will be shown in the weigher screens and will show information about;
Serial number, number of calibration certificates, the minimum allowed weight value and
the maximum allowed weight value.
This text can only be changed by an authorized person.
Class:
The class is only used when the weigher is certified and sealed by an authorized person
and depends on the used load cells and the construction.
The class will be shown in the weigher screens.
The class can only be changed by an authorized person.
Indicator set up 05
Page 2-11