FLEX 2100
19
Indicator set up 03
Operation Mode:
This parameter defines the operation mode of the unit.
The operation mode can be set at “Industrial” or Certified mode.
In industrial mode its always possible to change the indicator parameters and calibration
settings.
In certified Mode the unit will be sealed by marks and also the weighing parameters will be
blocked to comply with calibration laws.
If certified mode is selected, the zero band = 4% (+2 and -2%)
Also zero suppressing is disabled.
A weighing unit must be certified when its used for
Note:
measuring for trade aims.
Example: The normal step size is per 1, Range =100 and Step=50.
0-100
1
100-200
2
200-500
5
500-1000
10
1000-2000
20
2000-5000+
50
The maximum preset tare only works, and this is also the biggest value, to the first level of the
autorange when the indicator is set to certified. In this example the preset tare is valid to 100.
Indicator display
Step size per
Mode:
The multi interval function has 2 modes, “Multi range” en “Multi interval”.
With “Multi range” the highest shown stepsize will be reset after the signal has been lower
or equal to “0”.
With “Multi interval” the highest shown stepsize will be lowered when the signal reaches a
lower range.