1020 FMD
87
Entry Number
Enter the number of the desired record and conform with Enter or the Apply button. The record
is shown.
Record
The type of record. A record can be a
Header
or a
Data
record. A header records is the title of
the record and shows a tag, date and time. A header has several data records that show the
code, value and unit of the stored value.
Tag/Code
Tag shows a tag the header record belongs to. By default this is “Alibi 001” and cannot be
changed. Code shows the type of value stored in the data record.
Date/Value
Date shows the date of the header record. Value shows the stored measurement of the data
record.
Time/Unit
Time shows the time of the header record. Unit shows the stored unit of the data record.
UID
Every record has a Unique ID number, the UID.
This will print the total alibi memory to a connected printer. On the device it’s possible to print
only a part of the alibi memory. This is explained further on.
Clear
This will clear the total alibi memory.
When the print button is pressed, 6 records are added to the alibi memory. This is 1 header
record and 5 data records. Example:
Number
Record
Tag/Code
Date/Value
Time/Unit
UID
1
Header
Alibi 001
14-08-14
14:40:58
3298435072
2
Data
Force
0.783
kN
1755848705
3
Data
Peak
0.990
kN
2859737091
4
Data
Valley
0.129
kN
2346983429
5
Data
Hold
0.649
kN
0545136647
6
Data
T.I.R.
0.861
kN
2634031113
Summary of Contents for 1020 FMD
Page 7: ...1020 FMD 7 1 Overview Option 1 Option 2...
Page 19: ...1020 FMD 19 RS232 communication RS422 communication with multiple devices...
Page 21: ...1020 FMD 21 Hold function Peak Valley and T I R function...
Page 99: ...1020 FMD 99...
Page 103: ...1020 FMD 103...
Page 106: ...1020 FMD 106 Appendix I Menu structure...
Page 107: ...1020 FMD 107...
Page 108: ...1020 FMD 108...