1020 FMD
58
Signal
The current voltage, from the load cell or force sensor, in millivolts.
ADC
The current ADC value.
Points
show the stored calibration points.
Up to 10 calibration points can be stored to realize a multi-point calibration.
Add/Replace
point is used to add a calibration point.
When a point is added with a value that already exists, the existing point will be replaced. When
a point is added with a new value, it will be stored as a new point.
Delete point
is used to delete a calibration point.
Select the number of the calibration point (see Points) and confirm with Enter or the Apply
button. The calibration point will be deleted.