At this screen the UP/DOWN arrows are used to select what function to program. Press the
SELECT/PROG button to program the function. The setting or digit of a setting that can be
changed with the UP/DOWN arrows blinks. If there is more than one digit in a setting, use the
RIGHT/LEFT arrows to scroll within the possible digit locations within a setting. Pressing the
SELECT/PROG button will confirm a change- UNTIL THIS IS DONE, A SETTING
CHANGE IS NOT CONFIRMED. Pressing the HOME or BACK button before confirming a
setting change with the SELECT/PROG button will escape from that program menu without
saving changes made before confirming the entry.
IT IS ONLY RECOMMENDED THAT THE ITEMS IN SHOWN IN BOLD BELOW
ARE CHANGED. SCROLL TO SEE ALL SETTINGS:
Channel 3 Input Programming
> Port Type In 0-20mA
Measure Units YYY
Time Base Scalar
Decimal Point xx.x
Low Signal 4.000 mA
Low Units 0.0 YYY
High Signal 20.000 mA
High Units 70.0 YYY
Lo PV Alarm 0.00 YYY
Hi PV Alarm 0.00 YYY
Alarm Valid 000 sec
Time Alarm 0000 hrs
Comm Port Off
Log Type Off
Channel Name
Note: If the Alarm Points are set to 0.00 then there will be no alarms
After 30 seconds of inactivity in the program menu, the selected setting to change stops
blinking and the value before the setting was selected is restored.
Measure Units- This determines the label for the analog input units
Changing the Low and High Units- This determines the scale of the calculated units for the
analog input. The PV is determined by the following formula based on the Low and High
Units entered:
PV = ((PVmA Signal – 4) /16) * (High Units - Low Units) YYY
Changing the Alarm Point- Press the SELECT/PROG with the desired Alarm selected. Use
the RIGHT ARROW button on the keypad until you get to the digit that you want to change.
Use the DOWN ARROW/UP ARROW buttons on the keypad to change the values. To
change a digit from a numerical value to zero, change the value to a zero. When the correct
value has been selected, press the SELECT/PROG button and the value will be confirmed. To
escape from this menu without changing the value, press the BACK button and the main
channel menu will appear.