Input Programming
At this screen the UP/DOWN arrows are used to select what function to program. Press the
SELECT/PROG button to program the function. The setting or digit of a setting that can be
changed with the UP/DOWN arrows blinks. If there is more than one digit in a setting, use the
RIGHT/LEFT arrows to scroll within the possible digit locations within a setting. Pressing the
SELECT/PROG button will confirm a change- UNTIL THIS IS DONE, A SETTING
CHANGE IS NOT CONFIRMED. Pressing the HOME or BACK button before confirming a
setting change with the SELECT/PROG button will escape from that program menu without
saving changes made before confirming the entry.
Channel X Input Programming
> Lo PV Alarm 0.00 bar
Hi PV Alarm 0.00 bar
Alarm Valid 000 sec
Time Alarm 0000 hrs
Comm Port Off
PV Filter Off
Log Type Off
Channel Name
After 30 seconds of inactivity in the program menu, the selected setting to change stops
blinking and the value before the setting was selected is restored.
Changing the Low Pressure Alarm Point- Press the SELECT/PROG with Low Alarm selected
Use the RIGHT ARROW button on the keypad until you get to the digit that you want to
change. Use the DOWN ARROW/UP ARROW buttons on the keypad to change the values.
Press the RIGHT ARROW then UP ARROW to get the minus sign. To change a digit from a
numerical value to zero, change the value to a zero. When the correct value has been selected,
press the SELECT/PROG button and the value will be confirmed. To escape from this menu
without changing the value, press the BACK button and the main channel menu will appear.
Changing the High Pressure Alarm Point- Press the SELECT/PROG with High Alarm selected
Use the RIGHT ARROW button on the keypad until you get to the digit that you want to
change. Use the DOWN ARROW/UP ARROW buttons on the keypad to change the values.
Use the DOWN ARROW/UP ARROW buttons on the keypad to change the values. Press the
RIGHT ARROW then UP ARROW to get the minus sign. To change a digit from a numerical
value to zero, change the value to a zero. When the correct value has been selected, press the
SELECT/PROG button and the value will be confirmed. To escape from this menu without
changing the value, press the BACK button and the main channel menu will appear.
Note: If the High or Low Pressure Alarm Points are set to 0.00 then there will be no
alarm setting and an alarm will only occur if a sensor is disconnected