5.1.2. Displayed icons (symbols) and messages in LOOP/PFC function
Indicates the correct input terminal connectivity. The user should
connect the test leads to the appropriate terminals.
Indicates L connection is connected on the N input terminal and
vice-versa
Indicates no connection on the PE input terminal
If the wiring condition is other than normal, the Tester is limited on its measurements that can be
performed.
Notes:
1) Will not detect two hot wires in a circuit.
2) Will not detect a combination of defects.
3) Will not detect reversal of grounded and grounding conductors.
Indicates the battery status
100%
80%
50%
30%
Low Battery
Indicates high temperature and therefore cannot make any
measurements
Measuring
Function in use - measurement being carried out
RCD Trip
During the measurement, the RCD has tripped therefore no test
result obtained
-Noise-
Appears during the No Trip Loop Measurement, and indicates that
the displayed value may
not be accurate due to “mains” interference - test to be repeated
-66-