Compressor Replacement
Ensure that Generator is in Safely Isolated Condition
Removing the Compressor
1.
Disconnect the quick release coupling on the compressor hose, located
behind the compressor. Un-clip safety latch on electrical plug and
disconnect electrical power supply plug.
2.
Disconnect 10mm tubing from compressor intake 10mm push-fit fitting.
3.
Remove 10mm compressor mount nut from all four corners of the
compressor footplate.
4.
The compressor can now be lifted out.
Re-fitting the Compressor
1.
Lift the compressor into position and locate the compressor mounts
through their corresponding holes in the footplate.
2.
Replace the 10mm compressor mount nuts, and tighten, ensuring
mounts are not twisted.
3.
Reconnect the 10mm tubing from compressor intake to the push-fit
fitting.
4.
Reconnect the compressor hose quick release coupling
5.
Reconnect the electrical plug ensuring that the safety latch is clipped
securely back into position.