Principle of Operation
System Description
Normal Operation
The Corona Compressed Air Generator is designed specifically to minimize
operator involvement. Given that the system is installed correctly and is
serviced in accordance with the specified maintenance recommendations (see
Service Requirements), it should simply be a matter of turning the generator on
when required.
The generator has been factory set to produce the required outlet pressure.
Flow from the generator is unrestricted internally but is instead controlled by
the consuming nitrogen generator. This ensures minimal pressure drops
between generators. It is therefore important to note that normal operation of
the air compressor system can only be achieved when it is connected to the
nitrogen generator. Unconnected it will not reach pressure.
Connected as required, the compressor will periodically stop to eject moisture
from its drain bowls. Other than during this time, the compressor will run
continuously when powered on.
Figure 1: System Outline