Contact Peak Scientific Technical Help.
Yellow Service LED on front panel
is flashing.
•
Compressor is overdue for service.
•
Replace compressor – refer to Compressor Replacement OR
•
Refit compressor
•
Contact Peak Scientific Technical Help.
Yellow Pressure LED in front panel
is flashing.
•
LOW pressure has been detected for a minimum of 30mins
•
The compressor has not cycled.
•
Ensure ambient temperatures are within specification.
•
Check for leaks between the Generator and application.
•
Contact Peak Scientific Technical Help.