Corona CA 

Service Manual 

Page 23  

 

Trouble Shooting 

Problem 

Possible solution 

The Generator will not switch on 
and the power switch does not 
illuminate. 

 

Ensure power cord is plugged into the Generator 
and that the power socket is turned on. 

 

Check the fuse in the power cord plug. 

 

Check for any disconnected terminals at 
power socket connected at inside rear of unit. 

 

Check for loose connections at inside of power switch 
located on front of Generator. 

 

Check for 230VAC  going into DC power supply 

 

Check for 24VDC  coming from power supply 

 

Check circuit breakers are ON – switched up 

 

Contact Peak Scientific Technical Help.

 

The Generator will not switch on 
but the power switch is 
illuminated. 

 

Disconnect power cord from the rear of the 
Generator and open LH panel:

 

 

 

Check that both circuit breakers are turned on 
(switch in the up position). Reconnect power cord. 

 

Check that both compressor electrical plugs are connected. 

 

Check compressor contactors for loose wires. 

 

Check PLC for loose wires. 

 

Check pressure switch for loose wires. 

 

Connect power cord and power up Generator and open  LH 
panel: 

 

Check for 24V coming out of power supply. 

 

Check pressure switch LED is lit. 

 

Check PLC to ensure RUN light is illuminated. 

 

Check continuity across compressor contactors 

 

Contact Peak Scientific Technical Help.

 

Compressor is running but 
pressure is not building in system. 

 

Check drain outlet at rear of Generator is not 
passing air. 

 

Open the right hand panel and check if tank 
pressure gauge is rising thus pressure being built in receiver 
tanks. 

 

Open right hand panel and check pressure regulator. 

 

Check for leaks in system. 

 

Contact Peak Scientific Technical Help.

 

The mass spec is reporting low 
pressure. 

 

Check pressure gauge is set to factory settings.  

 

Check output flow from the Generator is set to factory 
settings.  

 

Check for leaks between receiver tanks and Generator 
output. 

 

Check for leaks between Generator output and Mass 
Spectrometer. 

 

Contact Peak Scientific Technical Help.

 

Yellow Service LED on front panel 
is on constantly. 

 

Compressor due for service.  

 

Replace compressor – refer to Compressor Replacement OR 

 

Refit compressor 

 

Summary of Contents for Corona CA

Page 1: ...Corona Compressed Air Service Manual 2019 Peak Scientific Rev 6 07 11 2019...

Page 2: ...on 7 System Description 7 Normal Operation 7 Service Requirements 8 Service Schedule 8 Product Overview 9 Electrical Panel Layout 10 Pressure switch overview 11 Air pressure switch settings 11 Compres...

Page 3: ...ically competent end users It has been written as a reference document where you can skip to the relevant information This manual does not cover all information that is included in other manuals Pleas...

Page 4: ...ions The attention of UK users is drawn to the Health and Safety at Work Act 1974 and the Institute of Electrical Engineers regulations A WARNING notice denotes a hazard It calls attention to an opera...

Page 5: ...lead to personal injury or injury to others and even death Isolating the Generator a Switch off the unit b Unplug the unit from the mains supply and remove the power cord from the rear of the unit c...

Page 6: ...Gas Outlets 1 Drain Outlets 1 Pressure Gauges 1 Electrical Requirements Voltage 100 100 115VAC 220 240 208 230VAC 230 VAC 10 Frequency 50 60 Hz 12x10x29 12x10x29 50 60 Hz Current 5 0 5 2A TBC TBC 2 3...

Page 7: ...been factory set to produce the required outlet pressure Flow from the generator is unrestricted internally but is instead controlled by the consuming nitrogen generator This ensures minimal pressure...

Page 8: ...Annual Service Kit www peakscientific com ordering 1 Table 1 Service schedule Compressors can be re fitted as an alternative to replacement up to a maximum of 3 times this is a more cost effective so...

Page 9: ...erview Tank Electrical Panel BPRV Re Heat Coil Compressor Pressure Regulator Drain Valves Silencer IEC Power Inlet Air Output and Drain Inlet Filter Dryer Membrane Duplex Filter Assembly Fin Tube Cool...

Page 10: ...Corona CA Service Manual Page 10 Electrical Panel Layout...

Page 11: ...Calibration is not required Air pressure switch settings Output Mode Set point SP Hysteresis HY Switching A default 8 3 1 6 NO default B default 9 5 0 7 NO default Table 2 Air pressure switch setting...

Page 12: ...e compressor contactors can be manually activated by sliding the switch on the front of the contactor to the left Valve Relay Accessing the electrical panel without safely isolating the Generator shou...

Page 13: ...lockwise and pulling upwards 2 The filter element is fitted into the filter lid Pull it out and clean the inside of the lid before fitting a replacement filter element into the lid 3 Clean the filter...

Page 14: ...r is in Safely Isolated Condition 1 Remove 6mm drain tubing from push fit stud situated on under side of both of the bowls 2 Remove the filter bowls by turning the bowl anti clockwise 3 Remove the fil...

Page 15: ...emove 10mm compressor mount nut from all four corners of the compressor footplate 4 The compressor can now be lifted out Re fitting the Compressor 1 Lift the compressor into position and locate the co...

Page 16: ...Corona CA Service Manual Page 16 Corona Compressed Air Pneumatic Schematic...

Page 17: ...Corona CA Service Manual Page 17 Corona Compressed Air Electrical Schematic 110V...

Page 18: ...Corona CA Service Manual Page 18...

Page 19: ...Corona CA Service Manual Page 19...

Page 20: ...Corona CA Service Manual Page 20 Corona Compressed Air Electrical Schematic 230V...

Page 21: ...Corona CA Service Manual Page 21...

Page 22: ...Corona CA Service Manual Page 22...

Page 23: ...res Check PLC for loose wires Check pressure switch for loose wires Connect power cord and power up Generator and open LH panel Check for 24V coming out of power supply Check pressure switch LED is li...

Page 24: ...refer to Compressor Replacement OR Refit compressor Contact Peak Scientific Technical Help Yellow Pressure LED in front panel is flashing LOW pressure has been detected for a minimum of 30mins The co...

Page 25: ...ne 2277 Zuchongzhi Road Pudong New Area Shanghai 201203 China Peak Scientific Mexico Solon 352 Col Los Morales Polanco 11530 Mexico D F Mexico Peak Scientific Australia PO Box 65 Belmont Victoria 3216...

Reviews: