![PCS PCS 4000 User Manual Download Page 18](http://html.mh-extra.com/html/pcs/pcs-4000/pcs-4000_user-manual_1516348018.webp)
18
USER GUIDE
PCS 4000
SINGLE WELL CONTROLLER
Plunger has surfaced, but Plunger Travel Time History
(READ 06) shows NO PLUNGER instead of a travel time
Make sure sensor is turned on: Press READ 09. The
controller’s display should show SENSOR: ON.
If the display shows SENSOR: ON, go to Test 1.
If the display shows SENSOR: OFF, press SET 09
and then press ON. Go to Test 2, step 2.
Test 1: Check controller by disconnecting the sensor.
Disconnect the sensor from the sensor and ground
1
terminals on the inside of the controller’s faceplate.
Press ON. The controller should count down the OPEN
2
TIME.
If the controller goes from OPEN TIME to DELAY
TIME (SALES), the controller’s circuit board is not
working properly. Call your PCS sales and service
representative.
If the controller does not go to DELAY TIME
(SALES), continue with the next step.
Use a wire to short the sensor and ground terminals.
3
The controller is OK if it goes to DELAY TIME (SALES).
Reconnect the sensor to the terminals. Make sure the
4
wires are put back in correctly.
Test 2: Check sensor.
Press ON.
1
2 Run a steel wrench along the back of the sensor.
If the controller goes to DELAY TIME (SALES), the
sensor is OK.
If the controller does not go to DELAY TIME
(SALES), call your PCS sales and service
representative.
Test 3: Check sensor alignment.
Realign the sensor by 90 degrees.
1
Example: If the sensor is aligned on the north or south
side, realign it on the east or west side.
If realigning the sensor does not solve the problem, call
2
your PCS sales and service representative.
Short-term fix for the sensor problem: Use the CASING
DROP PRESSURE feature (see page 12).
Plunger has surfaced, but Plunger Travel Time History
(READ 06) shows C instead of A or B
Make sure sensor is turned on: Press READ 09. The
controller’s display should show SENSOR: ON.
If the display shows SENSOR: ON, go to Test 1.
If the display shows SENSOR: OFF, press SET 09
and then press ON. Go to Test 3, step 2.
Test 1: Check CASING DROP PRESSURE.
Press READ 20.
1
Make sure the CASING DROP PRESSURE is high
2
enough that the plunger surfaces before the pressure
is reached. For instructions, see page 12.
If the CASING DROP PRESSURE is set correctly, go to
Test 2.
Test 2: Check controller by disconnecting the sensor.
Disconnect the sensor from the sensor and ground
1
terminals on the inside of the controller’s faceplate.
Press ON. The controller should count down the OPEN
2
TIME.
If the controller goes from OPEN TIME to DELAY
TIME (SALES), the controller’s circuit board is not
working properly. Call your PCS sales and service
representative.
If the controller does not go to DELAY TIME
(SALES), continue with the next step.
Use a wire to short the sensor and ground terminals.
3
The controller is OK if it goes to DELAY TIME (SALES).
Reconnect the sensor to the terminals. Make sure the
4
wires are put back in correctly.
Test 3: Check sensor.
Press ON.
1
2 Run a steel wrench along the back of the sensor.
If the controller goes to DELAY TIME (SALES), the
sensor is OK.
If the controller does not go to DELAY TIME
(SALES), call your PCS sales and service
representative.
Test 4: Check sensor alignment.
Realign the sensor by 90 degrees.
1
Example: If the sensor is aligned on the north or south
side, realign it on the east or west side.
If realigning the sensor does not solve the problem, call
2
your PCS sales and service representative.
Short-term fix for the sensor problem: Use the CASING
DROP PRESSURE feature (see page 12).