- PCME QAL 181 MANUAL: 76 -
Use the scroll bar to move the cursor to the required channel.
Press Edit to change sensor to Maintenance Mode.
The following table details the status of various functions according to sensor status:
Sensor Status
Function
Enabled
Maint
Disabled
Emission Alarms
(Control Unit relays
and ROM devices)
Active
Inactive
Inactive
Comms and Self Test
Alarms
Active
Active
Inactive
4-20mA
and
AOM
Outputs
Active
Set to 0
Set to 0
Bar Graph Displays
Active
Active
(labelled MAINT)
Set to 0
Logging
Active
Active
(labelled MAINT)*
Inactive
Alarm Log
-
Maintenance
Alarm
-
*Data may be excluded from PCME Dust Tools (Dust Reporter) reports.
B
A
C
K
E
D
I
T
Stack 2 Dust
MAINT
Stack 3 Dust
MAINT
Stack 4 Dust
ENABLED
Stack 1 Dust
ENABLED
Stack 5 Dust
DISABLED
Summary of Contents for QAL 181 X
Page 2: ...PCME QAL 181 MANUAL 2...
Page 15: ...PCME QAL 181 MANUAL 15...
Page 26: ...PCME QAL 181 MANUAL 26 5 4 2 1 3...
Page 30: ...PCME QAL 181 MANUAL 30 Figure 3 3 PLUS System Daisy Chained...
Page 87: ...PCME QAL 181 MANUAL 87...