5

 

2.8  Setup the Min/Max function 

 

In 

HOLD

 mode, press the mode button repeatedly until enter into the Min/Max 

mode. Press the up/down button to setup which mode you want. 

MIN

 indicates the minimum value of current measurement; 

MAX

 indicates the 

maximum temperature value of current measurement.   

 

 

2.9  Setup high and low alarm 

 

1) High alarm 

In Hold mode, press the mode button 

  until enter the high alarm mode. Press 

the up/down button to set the alarm on or off. 

When the signal of “

” shows on the display, it means the high alarm is 

activated. 

After set high alarm, you can press the mode button to set the temperature alarm 

value. 

 

 

2) Low alarm   

In Hold mode, press the mode button 

  utill enter to the low alarm mode. 

Press the up/down button to set the alarm on or off.   

When the signal “

”shows on the display, it means the low alarm is activated. 

After set low temperature alarm, you can press the mode button to set the 

temperature alarm value. 

 

 

Summary of Contents for PCE-IR 1300

Page 1: ...2 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kin...

Page 2: ...the parameters 2 7 Setup emissivity 2 8 Setup the Min Max function 2 9 Setup high and low alarm 2 10 Setup the temperature unit 2 11 Setup offset 3 Data Logger 3 1 Storing data 3 2 Datalogger recall...

Page 3: ...you will enter to another function setup e g press one time to enter to emissivity setup press it again will save emissivity and enter to MIN MAX setup and so on Up Laser With this button you can setu...

Page 4: ...2 1 3 External function introduction 1 Trigger 2 Battery compartment cover 3 USB interface 4 Telescope 4 1 2 3...

Page 5: ...7 seconds if no button is pressed 2 4 Setup the laser Pull the trigger keep it pressed and then press up button to activate or deactivate the laser The laser symbol in the display only if the trigger...

Page 6: ...r modify the requested value by pressing the up and down buttons Press mode button to save the value and enter to next function If you have not activated any button for 7 seconds the unit will not sav...

Page 7: ...button until enter the high alarm mode Press the up down button to set the alarm on or off When the signal of shows on the display it means the high alarm is activated After set high alarm you can pr...

Page 8: ...unit selecting mode and press the up down button to select the requested temperature unit Centigrade scale Fahrenheit scale 2 11 Setup offset In Hold mode press the mode button repeatedly until enter...

Page 9: ...showing a disk and an icon of MAX appear above the current temperature result The small numbers indicate the memory position You can choose the memory position from 1 2000 with the up down buttons Pr...

Page 10: ...Clear the memory In the Storing Data mode choose the memory position to 0 by pressing the down button Press the mode button will clear the memory The three buzzer signals confirms the successful clea...

Page 11: ...start the SETUP EXE on the CD ROM and install the software follow the instructions of the wizard USB interface driver installation 1 Insert USB Cable into computer s USB port Computer will find the ne...

Page 12: ...oftware the communication will be displayed in the status bar of the windows Choosethe correct COM Port via MENU Setup interface There will be a signal of PCE IR 1000 connected show on the left bottom...

Page 13: ...ak values Manual scaling can be done at any time using the control elements of the temperature axis 4 Stop the measurement To stop the current measurement please press the stop key Menu measurement st...

Page 14: ...2 1 5 Specifications 5 1 Optical Specifications PCE IR 1300 D S 120 1 PCE IR 1800 PCE IR 1600 D S 300 1...

Page 15: ...0 100 1 000adjustable Aim mode Laser telescope aim Signal processing MAX MIN Scan Hold Function Alarm functions Audible a nd visible HIGH LOW alarm Data storage 2000 point Backlight Three color Outpu...

Page 16: ...100ms 95 Emissivity 0 100 1 000 adjustable Aim mode Laser telescope aim Signal processing MAX MIN Scan Hold Function Alarm functions Audible a nd visible HIGH LOW alarm Data storage 2000 point Backli...

Page 17: ...e 100ms 95 Emissivity 0 100 1 000 adjustable Aim mode Laser telescope aim Signal processing MAX MIN Scan Hold Function Alarm functions Audible a nd visible HIGH LOW alarm Data storage 2000 point Backl...

Page 18: ...issivity 8 14 m Aluminum oxidized 0 2 0 4 Asbestos 0 95 Asphalt 0 95 Basalt 0 7 Ceramic 0 95 Concrete 0 95 Copper oxidized 0 4 0 8 Fabric 0 95 Glass plate 0 85 Gold 0 01 0 1 Gravel 0 95 Ice 0 98 Iron...