7. Operation

7.1 Operating and display elements 

Fig. 2 LC display 

The  indication  of  the  measured  value  as  well  as  the 
configuration  of  the  individual  parameters  occurs  menu-
driven  via  the  LC  display.  The  individual  functions  can  be 
set with the  help of three miniature push  buttons  located in 
the front. 
The menu system is a closed system allowing you to scroll 
both  forward  and  backward  through  the  individual  set-up 
menus  to  navigate  to  the  desired  setting  item.  All  settings 
are  permanently  stored  in  an  EEPROM  and  therefore 
available again even after a battery change. 

7.2 Structure of the menu system

 

8. Maintenance

In  principle,  this  device  is  maintenance-free.  If  desired,  the 
housing  of  the  device  can  be  cleaned  when  switched  of 
using a damp cloth and non-aggressive cleaning solutions. 

With certain media, however, the diaphragm may be pollut-
ed  or  coated  with  deposit.  It  is  recommended  to  define 
corresponding service intervals for control.. After placing the 
device  out  of  service  correctly,  the  diaphragm  can  usually 
be cleaned carefully with a non-aggressive cleaning solution 
and a soft brush or sponge. If the diaphragm is calcified, it is 
recommended  to  send  the  device  to  PCE  Instruments  for 
decalcification.  Please  read  therefore  the  chapter 

“Repair” 

below. 

!

A false cleaning of the device can cause an irreparable 
damage on the diaphragm. Therefore never use point-
ed objects or pressured air for cleaning the diaphragm. 

9. Return

Before every return of your device, whether for recalibration, 
decalcification,  modifications  or  repair, it has  to  be  cleaned 
carefully and packed shatter-proofed. You have to enclose a 
notice  of  return  with  detailed  defect  description  when 
sending  the  device.  This  form  can  be  provided  by  PCE 
Instruments.  If  your  device  came  in  contact  with  harmful 
substances, a declaration of decontamination is additionally 
required.    Should  you  dispatch  a  device  without  a 
declaration  of  decontamination  and  if  there  are  any  doubts 
in  our  service  department  regarding  the  used  medium, 
repair  will  not  be  started  until  an  acceptable  declaration  is 
sent. 

If  the  device  came  in  contact  with  hazardous 
substances,  certain  precautions  have  to  be 
complied with for purification! 

10. Warranty conditions

The  warranty  conditions  are  subject  to  the  legal  warranty 
period  of  24  months  from  the  date  of  delivery.  In  case  of 
improper use, modifications of or damages to the device, we 
do  not  accept  warranty  claims.  Damaged  diaphragms  will 
also  not  be  accepted.  Furthermore,  defects  due  to  normal 
wear are not subject to warranty services. 

11. Declaration of conformity / CE

The  delivered  device  fulfils  all  legal  requirements.  The 
applied  directives,  harmonised  standards  and  documents 
are listed in the EC declaration of conformity. 
Additionally, the operational safety is confirmed by the CE 
sign on the manufacturing label. 

12. Menu list

 

 

 

 

 

 

 

 

 

 

 

 

13. Disposal

The  device  must  be  disposed  according  to  the 
European 

Directives 

2002/96/EG 

and 

2003/108/EG (on waste electrical and electronic 
equipment)  Waste  of  electrical  and  electronic 
equipment  may  not  be  disposed  by  domestic 
refuse! 

WARNING!  Depending  on  the  measuring  medium, 
deposit  on  the  device  may  cause  danger  for  the  user 
and 

the 

environment. 

Comply 

with 

adequate 

precautions for purification and dispose of it properly. 

limit exceeding 

neg. / pos. 

symbol for 

indicate low 

batteries 

unit 

measured value 

1 P MIN 

Minimum pressure display 

▼–button: puts the current pressure as minimum value
▲–button: puts the value on zero

2 P MAX 

Maximum pressure display 

▼–button: puts the current pressure as maximum value
▲–button: puts the value on zero

3 LOWER

 

Displaying of the lower range  

This value was determined on the order and cannot be changed. 

4 UPPER

 

Displaying of the upper range 

This value was determined on the order and cannot be changed.

 

5 DECIMA

 

Setting of the decimal point position 

 

Depending on the nominal pressure range and on the set unit, only a limited number of positions after 
the decimal point can be displayed. 

6 UNIT

 

Setting of the pressure unit 

Permissible units: bar, mbar, PSI, InHg, cmHg, mmHg, hPa, kPa, Mpa, mH

2

O, InH

2

O. 

 

Along with the unit, the decimal point position has probably to be changed in order to get a correct 
indication of the measured value. Besides, depending on the nominal pressure range, perhaps not all 
available units can be used.  

7 SW OFF

 

Configuration of the switch-off automatic 

Meaning of the permissible number: 
"0": switch-off automatic is turned off 
"1" 

– "5": automatic switch-off in 1 to 5 minutes 

8 CAL ZP

 

Calibration of initial point 

If you detect a shifting of the measured value deviating from the offset, the display can be re-calibrated. For 
this, a pressure reference is necessary if the offset differs from the ambient pressure. The used pressure 
must be identical to the starting point of the pressure measuring range. For reading the new pressure into 
the device, push the ▲ button. 
Please note the following diviation: 
-

1 … x bar: the offset is calibrated at -0.9 bar; during calibration we check whether the device will be within 

tolerance at -1 bar (in theory); for re-calibration a pressure reference of -0.9 bar is necessary

 

0 … x bar abs.: the offset is calibrated at 0.1 bar abs.; during calibration we check whether the device will 
be within tolerance at 0 bar abs.; for re-calibration a pressure reference of 0.1 bar is necessary 

 

If the re-calibration leads to a worsening of the original calibration, e. g. as a result of a defect pressure 
reference, the defaults can be re-set by the menu item "LD FAC" according to your order 

9 CAL EP

 

Calibration of end point  

If you detect a shifting of the measured value deviating from the end point, the display can be re-calibrated. 
For this, a pressure reference is necessary if the offset differs from the ambient pressure. The used pres-
sure must be identical to the end point of the pressure measuring range. For reading the new pressure into 
the device, push the ▲ button. 

 

If the re-calibration leads to a worsening of the original calibration, e. g. as a result of a defect pressure 
reference, the defaults can be re-set by the menu item "LD FAC" according to your order.

 

10 LD FAC

 

Load defaults  

To load the defaults you have to push the ▲–button. After the action "LOADED" and "OK" appers in the 
display for a short time. The configuration mode will be left automatically. 

-

▲ / on button: 

 with this button you turn the device on; in the operating mode you move forward in the menu system or 
increase the displayed value 

-

▼ / off button:

   with this button you turn the device off; in the operating mode you move backwards in the menu system or 

decrease the displayed value 

-

OK-button:

 

with this button you enter the operating mode; besides, it is used to activate the different menu items and 
to confirm the set values 

To configure the different menu items, set the desired values by pushing the "▼"  or "▲" buttons. Confirm the setting with the 
„OK“ button and the menu item will start blinking to indicate that you can start the configuration. 
To save the configured values or to leave a menu item, you also have to push the “OK” button. 

 

Changes of the adjustable parameters become only effective after pushing the OK button and leaving the menu item. After 
leaving the menu system, all parameters will be checked against each other and in reference to the characteristics of the 
device. If the message "OK" appears in the display for some seconds, the configuration was successfully. If the message 
"ERROR" appears, at least one of the set values is out of the permissible range. For example, an error will occur if a device 
with a nominal pressure range of 400 bar should be set on 4 positions after the decimal point. If an error has been detect-
ed, the lastly runnable parameters will be set again.

 

 

After configuring the  unit, the conversion  of the  pressure range (in menus UPPER  and LOWER) into the  new  unit will only 
occur after leaving the complete menu system. Besides, depending on the number of displayed figures of the respective nom-
inal pressure range, probably not all available units (in menu UNIT) can be used.

 