PCE Instruments PCE-BSW 3 User Manual Download Page 9

 

 

©  PCE Instruments 

 

Charging the battery  

If  the  battery  voltage  is  low,  the  light  next  to  "UnderVoltage"  on  the  left  of  the  display  glows. 
Connect the mains adaptor and charge the battery. The connector is located at the bottom of the 
housing, in the recess in front of the front right foot of the balance. During charging, the light next 
to "InCharge" to the left of the display glows. 

Troubleshooting 

Problem 

Possible cause 

Solution 

No display 

Low or no voltage 

> Recharge batteries in time 
> Check whether charging via 
   the mains adaptor works 

Weight display 
not stable 

- Influence by the environment 
  (vibration, 
  temperature fluctuations), 
- Objects between weighing plate 
  and housing  
- The weighing material is not stable 

> Keep the environmental 
   conditions as stable as 
   possible 
> Remove objects 
> Fix the weighing product  

Weighing is not 
correct 

- No calibration 
- No taring before weighing net 
  weights 
- Readability, decimal places, unit 
  set incorrectly 
- With very small values, the display 
  remains 0 if the "E-0" value is too 
  high or automatic zero tracking has 
  been activated 

> Calibrate and -if necessary- 
    adjust 
> Tare before weighing 
> Check the set values 
> For small quantities, make 
   sure that E-01 and t0-OFF 
   have not been changed 

"CE-Err" 

Scale displays  
"CE-Err" when switched on 

> Reconnect the mains adaptor 
   to charge the battery 

"-Err-" 

- Quick calibration with wrong weight  
- Upper limit value outside the  
  weighing range of the balance 

> Use suitable calibration 
   weight 
>Adjust limit value 

„- - - - - -“ 

Weight of the item to be weighed 
exceeds the weighing range. 

Only load up to the specified 
max. weight 

 
 
 

 

Summary of Contents for PCE-BSW 3

Page 1: ...CE BSW 3 and PCE BSW 7 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last...

Page 2: ...al specifications 2 3 Factory settings 2 4 Delivery scope 2 5 Device description 3 6 First use of the balance 4 7 Function settings 4 7 1 Notes on changing the function settings 6 8 Charging the batte...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...s 287 x 271 x 127 mm 11 3 x 10 7 x 5 in Total weight 4 5 kg Interfaces Mains adaptor Output DC 9V 200 0mA outside minus inside plus 3 Factory settings PCE BSW 3 PCE BSW 7 Weighing range 3 000 g 7 500...

Page 5: ...imit 7 lb kg g display of the set weight unit 8 Display 9 Net Weight net weight lights up when tared Cancel taring with 0 or T key 10 Zero lights up when zero is displayed 11 ON OFF key 12 Key for fun...

Page 6: ...dC 6 6 0 0 o Current programme version number Weighing range in kg Current voltage CE Err means the mains adaptor is connected Current weight on the weighing platform Power saving mode after 1 minute...

Page 7: ...dash line appears then the measured value and the calibration is finished F 3 Dual range balance not available FUNC 1 FUNC 2 Dual range not available T to change MODE to continue to F 4 F 4 AD convert...

Page 8: ...nown weight whether the balance displays correctly Otherwise you must adjust the balance see item 1 5 Changing the weight units You can select grammes g kilogrammes kg pounds lb and carats CT If one o...

Page 9: ...weighing material is not stable Keep the environmental conditions as stable as possible Remove objects Fix the weighing product Weighing is not correct No calibration No taring before weighing net wei...

Page 10: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 11: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 058...

Reviews: