PCE Instruments PCE-BSW 3 User Manual Download Page 8

 

 

 

©  PCE Instruments 

 

7.1 

Notes on changing the function settings  

1. General 
If you change default settings, check the display values afterwards with a known weight. If there 
are deviations, recalibrate/adjust the balance. To do this, call up single-point (F---2) or three-point 
calibration (F---5) in the function menu. For the one-point calibration you can select the weight 
yourself. For three-point calibration, fixed weights are required. During calibration, the balance is 
automatically adjusted to the respective weights and ambient conditions. 
 
2. Changing the weighing range (

Should not be done by the customer!)

 

After each change of the weighing range, the balance should be adjusted (see item 1). 
Please do not change the weighing range as the load cells are not designed for this

.  

 
3. Changing readability/resolution (

Should not be done by the customer!)

 

The possible settings are

 

01 / 02 / 05 / 10.

 

 
4. Changing decimal places (

Should not be done by the customer!)

 

You can set 0, 1 or 3 decimal places. It is possible to set values here that do not match the other 
settings  (e.  g.  three  decimal  places  for  the  unit  gramme,  no  decimal  places  for  the  unit 
kilogramme). After changing the decimal places, check with a known weight whether the balance 
displays correctly. Otherwise, you must adjust the balance (see item 1). 

 
5. Changing the weight units 
You can select grammes (g), kilogrammes (kg), pounds (lb) and carats (CT). If one of the first 
three units is selected, the respective indicator on the right below the display glows. 
After  changing  the  weight  unit,  check  with  a  known  weight  whether  the  balances  displays 
correctly. Otherwise, you must adjust the balance (see item 1). When kg is set, the decimal place 
changes with it. This must be adjusted again when changing to grammes. 
 
6. Changing the sensitivity 
Here, you can set how sensitively the balance reacts at very low loads. 
For example, if you set the sensitivity to E-03, the balance will only display from 0.3 g when the 
resolution is 0.1 g and from 3 g when the resolution is 1 g. If you want the balance to display even 
the smallest quantities, set the sensitivity to E-01.  
 
7. Limit value function - P-OFF/ P-ON 
Here, you can activate that a visual indication appears on the left  below the display if upper or 
lower weight limits are not met. If the function is activated, you can set the weight limits within the 
weighing range under F---6.  
 
8. Zero tracking 
With  automatic  zero  tracking,  very  small  quantities  are  not  added  up  but  ignored  in  order  to 
stabilise the zero point. This "zeroing" of very small quantities can cause undesired false readings 
in some cases. 

 
 
 

 

Summary of Contents for PCE-BSW 3

Page 1: ...CE BSW 3 and PCE BSW 7 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last...

Page 2: ...al specifications 2 3 Factory settings 2 4 Delivery scope 2 5 Device description 3 6 First use of the balance 4 7 Function settings 4 7 1 Notes on changing the function settings 6 8 Charging the batte...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...s 287 x 271 x 127 mm 11 3 x 10 7 x 5 in Total weight 4 5 kg Interfaces Mains adaptor Output DC 9V 200 0mA outside minus inside plus 3 Factory settings PCE BSW 3 PCE BSW 7 Weighing range 3 000 g 7 500...

Page 5: ...imit 7 lb kg g display of the set weight unit 8 Display 9 Net Weight net weight lights up when tared Cancel taring with 0 or T key 10 Zero lights up when zero is displayed 11 ON OFF key 12 Key for fun...

Page 6: ...dC 6 6 0 0 o Current programme version number Weighing range in kg Current voltage CE Err means the mains adaptor is connected Current weight on the weighing platform Power saving mode after 1 minute...

Page 7: ...dash line appears then the measured value and the calibration is finished F 3 Dual range balance not available FUNC 1 FUNC 2 Dual range not available T to change MODE to continue to F 4 F 4 AD convert...

Page 8: ...nown weight whether the balance displays correctly Otherwise you must adjust the balance see item 1 5 Changing the weight units You can select grammes g kilogrammes kg pounds lb and carats CT If one o...

Page 9: ...weighing material is not stable Keep the environmental conditions as stable as possible Remove objects Fix the weighing product Weighing is not correct No calibration No taring before weighing net wei...

Page 10: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 11: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 058...