PCE Instruments PCE-BSW 3 User Manual Download Page 6

 

 

 

©  PCE Instruments 

 

First use of the balance  

Open the package carefully and keep the packaging for possible later shipment of the balance. 
Place the weighing plate on the balance and connect the mains adaptor. 
 
Set-up site 

1. 

Keep the set-up site clean and dry. 

2. 

Place the balance on a flat and stable surface. 

3. 

Avoid draughts and temperature fluctuations, e. g. through open windows and doors, 
installation under heaters or in direct sunlight. 

4. 

Avoid the effect of vibration on the balance. 

5. 

Make sure that the balance is not influenced by other devices. 

6. 

Static charge may occur in dry environments.  

7. 

Keep the balance away from objects with magnets and from magnetic fields. 

8. 

Do not use the balance in areas where there is a risk of explosion. 

 
To align the level horizontally, you can adjust the feet. The bubble of the spirit level should be 
exactly in the circle marked in red. 
 
 
When the balance is started, the display shows the following, one after the other: 
 

F5 

– 7.1 

C    

– 03 

dC - 6.6   
0,0 

- Current programme version number 
- Weighing range in kg 
- Current voltage (CE-Err means the mains adaptor is connected) 
- Current weight on the weighing platform 
- Power-saving mode (after 1 minute without load) / Press any key to return 
  to weighing mode. 
 

 
Setting the filter of the digital display 
If you press "MODE" and "-> T<-" in weighing mode, you can switch between three speeds (F ... 
high, M ... medium, L ... low). Pressing the key combination several times switches between the 
setting options. 
 

Function settings  

Press the "On/Off" key and then hold down the MODE key until "F---1" appears on the display. 
Press MODE to go to the basic settings, press >0< to go to F---2, F---3, F---4, etc. 
 

 

Function 

Description 

F---1 

Basic settings 

Continue to C03 with MODE, to F---2 with >0< 

C-03 

Weighing range 

1.5 / 3.0 / 6.0 / 7.5 / 15 / 30 in kg  
Select value with >T<, continue with MODE 

d-01 

Readability, resolution 

01 / 02 / 05 / 10 select with >T<, continue with MODE 

dP-1 

Decimal places 

Select 0 / 1 / 3 digits with >T<, continue with MODE 

UN-00 

Weight unit 

UN00 for g  
UN01 for kg  
UN-02 for lb  
UN-03 for CT       
Select with >T<, continue with MODE 

br-3 

Display brightness 

0 / 1 / 2 / 3 select with >T<, continue with MODE 

(br-0 stands for the highest brightness) 

Summary of Contents for PCE-BSW 3

Page 1: ...CE BSW 3 and PCE BSW 7 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last...

Page 2: ...al specifications 2 3 Factory settings 2 4 Delivery scope 2 5 Device description 3 6 First use of the balance 4 7 Function settings 4 7 1 Notes on changing the function settings 6 8 Charging the batte...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...s 287 x 271 x 127 mm 11 3 x 10 7 x 5 in Total weight 4 5 kg Interfaces Mains adaptor Output DC 9V 200 0mA outside minus inside plus 3 Factory settings PCE BSW 3 PCE BSW 7 Weighing range 3 000 g 7 500...

Page 5: ...imit 7 lb kg g display of the set weight unit 8 Display 9 Net Weight net weight lights up when tared Cancel taring with 0 or T key 10 Zero lights up when zero is displayed 11 ON OFF key 12 Key for fun...

Page 6: ...dC 6 6 0 0 o Current programme version number Weighing range in kg Current voltage CE Err means the mains adaptor is connected Current weight on the weighing platform Power saving mode after 1 minute...

Page 7: ...dash line appears then the measured value and the calibration is finished F 3 Dual range balance not available FUNC 1 FUNC 2 Dual range not available T to change MODE to continue to F 4 F 4 AD convert...

Page 8: ...nown weight whether the balance displays correctly Otherwise you must adjust the balance see item 1 5 Changing the weight units You can select grammes g kilogrammes kg pounds lb and carats CT If one o...

Page 9: ...weighing material is not stable Keep the environmental conditions as stable as possible Remove objects Fix the weighing product Weighing is not correct No calibration No taring before weighing net wei...

Page 10: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 11: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 058...