First use of the balance
Open the package carefully and keep the packaging for possible later shipment of the balance.
Place the weighing plate on the balance and connect the mains adaptor.
Set-up site
1.
Keep the set-up site clean and dry.
2.
Place the balance on a flat and stable surface.
3.
Avoid draughts and temperature fluctuations, e. g. through open windows and doors,
installation under heaters or in direct sunlight.
4.
Avoid the effect of vibration on the balance.
5.
Make sure that the balance is not influenced by other devices.
6.
Static charge may occur in dry environments.
7.
Keep the balance away from objects with magnets and from magnetic fields.
8.
Do not use the balance in areas where there is a risk of explosion.
To align the level horizontally, you can adjust the feet. The bubble of the spirit level should be
exactly in the circle marked in red.
When the balance is started, the display shows the following, one after the other:
F5
– 7.1
C
– 03
dC - 6.6
0,0
o
- Current programme version number
- Weighing range in kg
- Current voltage (CE-Err means the mains adaptor is connected)
- Current weight on the weighing platform
- Power-saving mode (after 1 minute without load) / Press any key to return
to weighing mode.
Setting the filter of the digital display
If you press "MODE" and "-> T<-" in weighing mode, you can switch between three speeds (F ...
high, M ... medium, L ... low). Pressing the key combination several times switches between the
setting options.
7
Function settings
Press the "On/Off" key and then hold down the MODE key until "F---1" appears on the display.
Press MODE to go to the basic settings, press >0< to go to F---2, F---3, F---4, etc.
Function
Description
F---1
Basic settings
Continue to C03 with MODE, to F---2 with >0<
C-03
Weighing range
1.5 / 3.0 / 6.0 / 7.5 / 15 / 30 in kg
Select value with >T<, continue with MODE
d-01
Readability, resolution
01 / 02 / 05 / 10 select with >T<, continue with MODE
dP-1
Decimal places
Select 0 / 1 / 3 digits with >T<, continue with MODE
UN-00
Weight unit
UN00 for g
UN01 for kg
UN-02 for lb
UN-03 for CT
Select with >T<, continue with MODE
br-3
Display brightness
0 / 1 / 2 / 3 select with >T<, continue with MODE
(br-0 stands for the highest brightness)