PCE Instruments PCE-BSW 3 User Manual Download Page 7

 

 

©  PCE Instruments 

 

PS-01 

Power saving mode 

PS-00 - no power saving  
PS-01 - power saving 
PS-02 - rear display off - select with >T< 

E

–01 

Sensitivity 

00 / 01 / 02 / 03 / 04 / 05 / 06 / 07 / 08 

P-OFF  

Activate weight limits 

Select P-OFF / P-ON   with >T<, continue with 
MODE 
Under F---6, the values can be set.  
When weighing, the following glows on the left below 
the display: 
- Down (red) - if value is below lower limit value 
- Pass (green) - if value is within specified range 
- Up (red) - if value is above upper limit value 

t0-OFF 

Zero tracking 

Automatically corrects deviations of the zero display 
- t0-OFF / t0-ON - select with >T< 
(should always remain activated) 

F---2 

Single-point calibration 

(We recommend using 
at least 2/3 of the max. 
load -PCE-BSW 3 - 2 kg 
and PCE-BSW 7 - 5 kg

When "LOAD" is displayed, place the weight on the 
weighing plate. Confirm with MODE. When 
"L0000.0" is displayed, enter weight with >0< and 
>T<. Confirm with MODE. A double dash line 
appears, then the measured value and the 
calibration is finished. 

F---3 

Dual range balance 

!! not available !!

 

"FUNC-1"/"FUNC-2" 

Dual range not available!!! 

>T< to change, MODE to continue to F---4 

F---4 

AD converter 

Display for checking the function of the load cell 

The numerical value must increase when the balance 
is loaded and decrease when the load is removed.  

F---5 

Three-point calibration 

(The required weights 
depend on the weighing 
range of the balance

). 

"CAL", then "

3000

" -> put on 3000 g 

Double line, then "

3000

" -> remove weight 

Double line, then "

2000

" -> load 2000 g.  

Double line, then "

2000

" -> remove weight 

Double line, then "

1000

" -> load 1000 g 

Double line, then "

1000

" -> remove weight 

Double line, then "0.0" -> calibration is finished 

F---6 

Set weight limits 

(for 

activation, see "P - 
OFF

")

 

Change position with >T<, change value with >0<,  
confirm entries with MODE, 
"L0000.0" lower, "H0000.0" upper limit value  

-END- 

End of settings menu 

Balance automatically goes to weighing mode 

 
Attention! 

The greyed-out menu items should not be used without consulting a service technician. 
Be sure to follow the instructions on the following page!  

 
 

 

Summary of Contents for PCE-BSW 3

Page 1: ...CE BSW 3 and PCE BSW 7 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last...

Page 2: ...al specifications 2 3 Factory settings 2 4 Delivery scope 2 5 Device description 3 6 First use of the balance 4 7 Function settings 4 7 1 Notes on changing the function settings 6 8 Charging the batte...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...s 287 x 271 x 127 mm 11 3 x 10 7 x 5 in Total weight 4 5 kg Interfaces Mains adaptor Output DC 9V 200 0mA outside minus inside plus 3 Factory settings PCE BSW 3 PCE BSW 7 Weighing range 3 000 g 7 500...

Page 5: ...imit 7 lb kg g display of the set weight unit 8 Display 9 Net Weight net weight lights up when tared Cancel taring with 0 or T key 10 Zero lights up when zero is displayed 11 ON OFF key 12 Key for fun...

Page 6: ...dC 6 6 0 0 o Current programme version number Weighing range in kg Current voltage CE Err means the mains adaptor is connected Current weight on the weighing platform Power saving mode after 1 minute...

Page 7: ...dash line appears then the measured value and the calibration is finished F 3 Dual range balance not available FUNC 1 FUNC 2 Dual range not available T to change MODE to continue to F 4 F 4 AD convert...

Page 8: ...nown weight whether the balance displays correctly Otherwise you must adjust the balance see item 1 5 Changing the weight units You can select grammes g kilogrammes kg pounds lb and carats CT If one o...

Page 9: ...weighing material is not stable Keep the environmental conditions as stable as possible Remove objects Fix the weighing product Weighing is not correct No calibration No taring before weighing net wei...

Page 10: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...