© PCE Instruments
5
PS-01
Power saving mode
PS-00 - no power saving
PS-01 - power saving
PS-02 - rear display off - select with >T<
E
–01
Sensitivity
00 / 01 / 02 / 03 / 04 / 05 / 06 / 07 / 08
P-OFF
Activate weight limits
Select P-OFF / P-ON with >T<, continue with
MODE
Under F---6, the values can be set.
When weighing, the following glows on the left below
the display:
- Down (red) - if value is below lower limit value
- Pass (green) - if value is within specified range
- Up (red) - if value is above upper limit value
t0-OFF
Zero tracking
Automatically corrects deviations of the zero display
- t0-OFF / t0-ON - select with >T<
(should always remain activated)
F---2
Single-point calibration
(We recommend using
at least 2/3 of the max.
load -PCE-BSW 3 - 2 kg
and PCE-BSW 7 - 5 kg
)
When "LOAD" is displayed, place the weight on the
weighing plate. Confirm with MODE. When
"L0000.0" is displayed, enter weight with >0< and
>T<. Confirm with MODE. A double dash line
appears, then the measured value and the
calibration is finished.
F---3
Dual range balance
!! not available !!
"FUNC-1"/"FUNC-2"
Dual range not available!!!
>T< to change, MODE to continue to F---4
F---4
AD converter
Display for checking the function of the load cell
The numerical value must increase when the balance
is loaded and decrease when the load is removed.
F---5
Three-point calibration
(The required weights
depend on the weighing
range of the balance
).
"CAL", then "
3000
" -> put on 3000 g
Double line, then "
3000
" -> remove weight
Double line, then "
2000
" -> load 2000 g.
Double line, then "
2000
" -> remove weight
Double line, then "
1000
" -> load 1000 g
Double line, then "
1000
" -> remove weight
Double line, then "0.0" -> calibration is finished
F---6
Set weight limits
(for
activation, see "P -
OFF
")
Change position with >T<, change value with >0<,
confirm entries with MODE,
"L0000.0" lower, "H0000.0" upper limit value
-END-
End of settings menu
Balance automatically goes to weighing mode
Attention!
The greyed-out menu items should not be used without consulting a service technician.
Be sure to follow the instructions on the following page!