USER MANUAL
____________________________________________________________________________________________________________________
13
12.
Internal calibration (autocalibration)
Internal calibration is performed automatically after each start-up, also in defined
time intervals and temperature changes.
To calibrate the balance with internal weight, simply empty the pan and press
key.
The beginning of internal calibration is signalised with –CAL– communicate on
the display.
Internal calibration settings (time intervals and temperature changes)
Press MENU to call functions and press
→
T
←
to choose calibration "CAL ..".
Calibration function options:
-CAL - calibration (see chapt. 11)
-C_LOAD – calibration weight
-C_tIME - time interval setting
-C_tEMP - temp. change setting
-rEPort - calibration report
Press
→
T
←
to choose option (to
accelerate press several times ) and
value.
MENU
MENU
T
T
T
T
T
0.0000g
0.0000g
0,1,2 ...
0.5, 1.0,...
2h
0.5 C
CAL ...
C_tIME
C_tEMP
C_tEMP
CAL
...
...
...