1.

 

To ensure proper weighing accuracy the balance is equipped with internal 
calibration system. The system automatically calibrates the balance every 2 
hours and with temperature changes more than 1°C (values modification 
possible) without user intervention. When the balance is moved to another 
localisation remember to level the balance and proceed with internal calibration.  

2.

 

Weighed sample should be placed in the centre of the pan. 

3.

 

In balances equipped with 

0

 

key and with d=e, make sure that zero 

indicator is displayed. If not, press 

0

 

key and wait until zero indication and 

zero indicator appears. In other balances the key does not operate. 

4.

 

The balance is equipped with a tare equal to its range. To tare the balance press 

T

 key (left or right). Storing a tare value does not extend measuring range, 

but only subtracts it from a load placed on the pan. To make weight control 
easier and to avoid range overdrawing, the balance is equipped with a load 
indicator (graduated in percentages). 

5.

 

Weighing result should be read when the indicator "

"  lights, which signalises 

stabilisation of a result. 

6.

 

When the balance is not used but it is necessary to be ready to work 
immediately, it can be switched off by pressing 

I

/

 key. The balance reading 

system is then switched off to "standby" mode (signalled by the indicator 
"OFF"). To switch the balance on press 

I

/

 key. The balance is immediately 

ready to operate maximum accuracy (after self tests). 

7.

 

The mechanism of the balance is a precise device, sensitive to mechanical 
strokes and shocks. 

8.

 

The balance should not be used to weigh ferrometic PCE-ABZ 200C materials 
due to accuracy decrease. 

 

Do not overload the balance more then 20% of maximum load 
(Max). Do not press the pan with a hand. 

Before transportation take off the pan (move it slightly and lift it 

up) and the pan base and preserve from any damages. 

Summary of Contents for PCE-ABZ 100C

Page 1: ...0 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton Unite...

Page 2: ...9 9 Operation principles 10 10 Testing and calibration 11 11 External calibration calibration options 11 12 Internal calibration autocalibration 13 13 Connecting a balance to a computer or a printer...

Page 3: ...ith an internal calibration system for proper accuracy control during operation The user should also own weight standard of OIML F2 or F1 class for periodical control of the balance weight value state...

Page 4: ..._________________________________________________________ 4 3 General view Front view 1 pan 2 support pan 4 display LCD 5 keyboard 6 legs 7 water level Rear view 7 6 1 2 8 9 4 5 3 ON OFF Adjust switch...

Page 5: ...weighed mass key 0 numeric key zeroing balances for direct sale use only key numeric key data output print transmission key numeric key internal calibration on and off key Menu numeric key Menu key I...

Page 6: ...city Max 100g 200g Min load Min 10mg 10mg Reading unit d 0 1mg 0 1mg Verification unit e 1mg 1mg Tare range 100g 200g Accuracy class I Working temperature 18 30 C Weighing time 8s Pan dimension 90mm B...

Page 7: ...propriate type supplied with the balance Pay attention that supply voltage is compatible with specified technical data Do not use the balance when its cover is opened Do not use the balance in explosi...

Page 8: ...the balance remember the following rules to ensure proper working conditions and user friendly operating setup the balance on an even flat surface leaving necessary room for easy access maintain prop...

Page 9: ...of the balance is in the middle 4 Place tin floor plate 4 and the pan ring 3 in the weighing chamber and gently put on the pan support 2 and finally the pan 1 5 Plug a feeder to the power socket at th...

Page 10: ...or right Storing a tare value does not extend measuring range but only subtracts it from a load placed on the pan To make weight control easier and to avoid range overdrawing the balance is equipped...

Page 11: ...balance a service centre should use calibration weight as stated in Technical Data table or of better accuracy with valid calibration certificate External calibration in verified balance should be pe...

Page 12: ...ess T to choose C_LOAD and calibration weight value Press T to choose CAL_EXt Wait until zeroing process is finished When LOAD communicate put a calibration weight on the pan Wait until calibration pr...

Page 13: ...and press key The beginning of internal calibration is signalised with CAL communicate on the display Internal calibration settings time intervals and temperature changes Press MENU to call functions...

Page 14: ...bel printers Serial port working modes Standard mode The balance sends weighing result after initialising signal from a computer or after pressing key Automatic mode when cooperating with a printer Th...

Page 15: ...0Dh Byte 16 LF 0Ah 2 Automatic mode After stabilisation of each weighing result excluding zero indication the balance sends 3 digit successive measurement number and a weighing result To clear the me...

Page 16: ...lance KAFKA printer KAFKA printer internal switches set up SW 1 SW 2 SW 3 SW 4 SW 5 SW 6 SW 7 SW 8 on off on off off on off off It is possible to use a transmission protocol other then LonG on custome...

Page 17: ...ial port working mode setting Other functions recipe ingredients summing density calculation and other special functions can be enabled as an option on customer request described in additional brochur...

Page 18: ...gend press a key balance indication forced change automatic change 16 Normal weighing When the pan is empty and indication is different than zero press T key Weighing result should be read when the in...

Page 19: ...cation if the pan is empty or if zero indication was acquired by pressing T key Function has options On function is activated OFF function is not activated CorrECt maximum zero flow to be automaticall...

Page 20: ...is advised that single piece weight is not less than one reading unit and sample weight used in first phase is bigger than 100 reading units To leave function press MENU key and then with T key chose...

Page 21: ...232CS Function alows to set RS232C parameters velocity bAud 1200 4800 9600 bits bit 7 8 parity control PArItY 0 1 Odd 0 1 continuous transmition without key 10 times per secund SEnd 0 1 Press MENU to...

Page 22: ...__________________________ 22 21 Date and time setting F dAt Function puts date and time into printout and alows to set actual Function has options dAtE date setting tIME time setting Changing value n...

Page 23: ...ibrate the balance contact nearest service 5 It is forbidden to make any repairs by unauthorised persons 6 To repair the balance please contact our nearest service Error communicates Communicate Possi...

Reviews: