background image

 

Manual 

 

14 

7.6.1 

Impact Device 

There are two kinds of impact device can be selected. Impact device D is for normal application and DL is 
for some specific situation, such as narrow step or deep blind hole. When you select impact device DL, 
you need to change impact body D to DL. (The impact body DL is optional, you need to order separately).  
Press ▼ to enter the Menu screen. Press ▼ or ▲ to select Impact device and press ► to enter. Press ▼ 
or ▲ again to select D or DL in cycle. Press ► to confirm and return to the settings menu, press and hold 
any key to save and exit to the measure screen. 

7.6.2 

Material selection 

Press ▼ to enter the Menu screen. Press ▼ or ▲ to select Material and press ► to enter. Press ▼ or ▲ 
again to select the material in cycle. Press ► to confirm after selection and return to the Menu screen, or 
press and hold any key to save and return to the Measure screen. The symbol of the chosen material will 
display at right upper of Measure screen. 
 
M1 

STEEL/CAST STEEL  M6 

CAST ALUMINUM 

M2 

ALLOY TOOL STEEL  M7 

BRASS 

M3 

STAINLESS STEEL 

M8 

BRONZE 

M4 

LAMELLAR IRON 

M9 

WROUGHT COPPER 

M5 

NODULAR IRON 

M10 

FORGING STEEL 

7.6.3 

Hardness scale conversion 

Press ▼ to enter the Menu screen. Press ▼ or ▲ to select Scale and press ► to enter. Press ▼ or ▲ 
again to select the hardness scale in cycle: HL→HRC→HRB→HB→HV→HS→HRA→SGM 
Press ► to confirm after selection and return to the Menu screen, or press and hold any key to save and 
return to the Measure screen. The symbol of the chosen unit will display at right of Measure screen. 

7.7 

No. of test 

Press ▼ to enter the Menu screen. Press ▼ or ▲ to select No. of Test option and press ► to enter, then 
you can view the current measurement counts for convenient maintenance of the instrument. Press ▼ or 
▲ to select Exit or Erase. Press ► to confirm. Select Erase to clear the measurement counts. 

7.8 

Language selection 

Press and hold ▼ to enter Language screen. Press ▼ and ▲ to select suitable language, press ► after 
selection to return to the measuring screen, or press and hold any key to exit without saving and return to 
the measuring screen. 

 

7.9 

Mean Time Setup 

Press ▼ to enter the Menu screen. Press ▼ or ▲ to select Mean time and press ► to enter. Press ▼ or 
▲ again to select the mean times in cycle:  
X(Off) →x=3→x=4→x=5→x=6→x=7. 
Press ► to confirm after selection and return to the Menu screen, or press and hold any key to save and 
return to the Measure screen. The symbol of the mean will display at bottom of Measure screen. For 
example, as x=3 is chosen, display x3-0. Display the mean value after three measurements are got.  

7.10  Upper / Lower Limit 

Press ▼ to enter the Menu screen. Press ▼ or ▲ to select Limits and press ► to enter. Press ▼ or ▲ 
again to select the upper/lower limit in cycle:  
X(Off) → (On) → Upper → Lower. Select “Off” to disable Upper/Lower limit alarm. Select “On” to enable 
Upper/Lower limit setting, and the audio alarm will be enabled automatically, prompting alarm sound after 
overrunning. Select “Upper” or “Lower” and press ► to enter limits setting screen. Press ▼ or ▲ to adjust 
value. Press and hold ▼ or ▲ for fast adjustment. Press ► to confirm after adjustments and return to 
Limits menu, select On and press ► save changes and return to the upper-level menu, or press and hold 

Summary of Contents for PCE-2500

Page 1: ...62 Fax 561 320 9176 info pce americas com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax...

Page 2: ...measurements brought about by coarse grinding or lathe scoring The roughness of the finished surface should not exceed 2 m 8 6 2 Requirements for the weight of the sample 8 6 3 Requirement for the su...

Page 3: ...e main body the tester with left hand while right hand is holding the loading tube 18 9 1 2 Push the loading tube with a little force against spring force toward tester until to lock the impact body 1...

Page 4: ...ip made of tungsten carbide is propelled against the sample surface by a spring force and then rebounds back At a distance of 1mm from the sample surface the impact and rebound velocity of the impact...

Page 5: ...s for the user or damage destruction of the device The device may only be used in the specified temperature humidity range Do not expose it to extreme temperatures direct sunlight extreme air humidity...

Page 6: ...curacy 2HL or 0 3 HL 800 Display Digital with high contrast OLED Impact direction Universal angle no need to setup impact direction Display mode Normal flip or upward downward Hardness scale HL HRC HR...

Page 7: ...A b N mm2 Leeb hardness value used with impact device D Leeb hardness value used with impact device DL Brinell hardness value Rockwell B hardness value Rockwell C hardness value Shore hardness value V...

Page 8: ...he surface of the base plate by moving it with a circular motion The direction of impact should be perpendicular to the coupling surface For the coupling operation the following prerequisites must be...

Page 9: ...o not move or flex during the test impact b Light weight samples should be rigidly coupled with a non yielding support such as a heavy base plate Clamping in a vice is of no value since the samples be...

Page 10: ...suring direction Upward key press it to move cursor upward Press and hold it to save the settings and exit Press it in measuring screen to switch in different display modes Press and hold it to delete...

Page 11: ...Manual 11 7 2 Diagram of operation...

Page 12: ...strument will be powered on The screen will display the measuring window after it is switched on you can start measurement At this moment all parameters will be performed by factory default new tester...

Page 13: ...r Exit Return to the Measure screen without modification Press to return to the Menu screen or press and hold any key to save and return to the Measure screen Default Restore factory settings Save as...

Page 14: ...at right of Measure screen 7 7 No of test Press to enter the Menu screen Press or to select No of Test option and press to enter then you can view the current measurement counts for convenient mainten...

Page 15: ...Only save the mean value Press to confirm and return to the Menu screen 7 11 3 MODE3 Manual Save Save the data by pressing after measuring Press to confirm and return to the Menu screen 7 11 4 View T...

Page 16: ...s After Calibration is done the symbol will display at upper of the screen 7 12 1 Calibration mode Press to enter the Menu screen Press or to select Calibration and press to enter Press to move to Mod...

Page 17: ...ation in calibration menu press to move to X then press to switch off the Calibration function and return to the Menu screen 7 12 5 Calibration for DL probe For DL probe you need to make calibration i...

Page 18: ...be and screw the guide tube onto the probe Tester equipped with DL impact device Note The block D supplied with the new tester indicates the standard HLD value instead of LDL value when measuring on s...

Page 19: ...eavy dust damp strong magnetic field and oil stain 10 1 Maintenance of the impact device The devices do not require any particular care other than periodic cleaning of the impact body and the guide tu...

Page 20: ...tive 2012 19 EU we take our devices back We either re use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instrumen...

Page 21: ...r fer PCE 2500 Ger teserie bis Dezember 2015 Version 1 2 25 02 2016 PCE Deutschland GmbH Im Langel 4 D 59872 Meschede Deutschland Tel 02903 976 99 0 Fax 02903 976 99 29 info pce instruments com www pc...

Page 22: ...arnhinweise 5 3 Spezifikationen 6 4 Ger tebeschreibung 6 5 Betriebsanleitung 9 5 1 Vorbereitende Ma nahmen 9 5 2 Bedienung des Ger ts 10 5 2 1 Einstellung der Men punkte 11 5 2 2 Wechsel des Schlagk r...

Page 23: ...rtemessung wurde erstmals 1978 eingesetzt Sie ist definiert durch das Verh ltnis von der R ckprallgeschwindigkeit eines einschlagenden K rpers zu dessen Einschlaggeschwindigkeit multipliziert mit dem...

Page 24: ...de wenn ein Schlagger t des Typs D verwendet wurde w re 35 9 HRCLD 35 H rtewert HL 9 H rtewert HRC L Leeb Methode D Schlagger t Verwendete Abk rzungen Abk rzung Bedeutung HLD Leeb H rtewert HB Brinell...

Page 25: ...s Geh use und die Messleitungen auf sichtbare Besch digungen berpr fen Sollte eine sichtbare Besch digung auftreten darf das Ger t nicht eingesetzt werden Weiterhin darf dieses Messger t nicht eingese...

Page 26: ...es Werkst cks 2 kg auf stabiler Auflage 50 g mit Koppelgel Mindestdicke des Werkst cks 3 mm mit Koppelpaste Datenspeicher 30 Dateien 100 Daten Messwert Material pro Datei Display Kontrastreiches digit...

Page 27: ...BETRIEBSANLEITUNG 7 Bedienschema Englisch...

Page 28: ...BETRIEBSANLEITUNG 8 Bedienschema Deutsch...

Page 29: ...erfahren eine Mindestdicke von f nf Millimetern nicht unterschreiten Anforderungen an Materialproben mit geh rteter Oberfl che Oberfl chengeh rtete St hle und insbesondere einsatzgeh rtete St hle erge...

Page 30: ...Inaktivit t nach 40 Sekunden aus Zum Einschalten schieben Sie das Laderohr in Richtung Ger t bis der Ausl seknopf gespannt ist Das Display wird aktiviert Nun ist das Ger t bereit zur Messung Die Eins...

Page 31: ...n Sie um zur ck in den Men bereich zu gelangen oder halten Sie eine beliebige Taste gedr ckt um zu speichern und zur ck in den Messmodus zu gelangen Standardeinstellungen auf Werkseinstellungen zur ck...

Page 32: ...LUMINIUMGUSS Umrechnung H rteskala Sie gelangen in den Men bereich indem Sie dr cken Dr cken Sie um Skala auszuw hlen und best tigen Sie mit Enter Verwenden Sie erneut die Tasten um eine der H rteskal...

Page 33: ...den Messmodus zur ck zu kehren Im unteren Bereich des Mess Displays wird das entsprechende Symbol angezeigt Wenn Sie z B x 3 ausw hlen wird Ihnen im Display x3 0 angezeigt was bedeutet dass der Mittel...

Page 34: ...kehren L schen L scht alle Daten der aktuellen Datei Wenn zum Beispiel Datei A0 angezeigt wird werden alle darin enthaltenen Daten gel scht Dr cken Sie damit Alle l schen angezeigt wird Wenn Sie mit b...

Page 35: ...ur HL Werte justiert werden w hrend im MODUS I Werte jeder Skala justiert werden k nnen Sobald Sie den Kalibriermodus deaktivieren werden nat rlich alle angezeigten Werte zur ckgesetzt es sei denn Sie...

Page 36: ...ist Nach erfolgter Kalibrierung sehen Sie das Symbol oben im Display Kalibrierung aus Wenn Sie die Kalibrierfunktion deaktivieren m chten dr cken Sie wenn Sie sich im Kalibriermen befinden die Taste u...

Page 37: ...in das entsprechende Men zu kommen Best tigen Sie die Sicherheitsabfrage mit oder kehren Sie durch Dr cken der Taste zur ck ins Funktionsmen Wenn Sie nicht innerhalb von einer Sekunde eine Option ausw...

Page 38: ...des Schlagk rpers Es bedarf keiner besonderen Wartung abgesehen von gelegentlicher nach ca 1000 2000 Messungen Reinigung des Schlagk rpers und des F hrungsrohres Schrauben Sie hierzu den St tzring ab...

Page 39: ...eben werden Der Endverbraucher ist zur R ckgabe gesetzlich verpflichtet Gebrauchte Batterien k nnen unter anderem bei eingerichteten R cknahmestellen oder bei der PCE Deutschland GmbH zur ckgegeben we...

Reviews: