4.2 

Keys  

Key 

Function 

MEAS / Hold 

- Switches the meter ON/OFF 
- Freezes the reading. Press the key again to continue 
  measurement. 
- Ends the calibration or adjustment and returns to measuring 
  mode  

CAL 

- Press the key to start the calibration. 
- Press and hold the key to open the setup menu. 
- In setting mode, press the key to select the default option. 

Enter 

Confirms the entries 

 

Measurement preparation  

5.1 

Insert batteries  

 

 

 

1. 

Unscrew the union nut of the electrode and pull the electrode off the measuring device, 
observing correct axial alignment. 

2. 

Insert two AAA batteries into the battery compartment. Observe correct polarity. 

3. 

Align  the  guide  groove  with  the  electrode  and  carefully  push  the  electrode  into  the 
meter. Screw the union nut of the electrode tight. 

5.2 

Preparation of the buffer solution  

Prepare 3 sealable containers with at least 250 ml content. Open the bags with the buffer powder 
and put the contents of one bag into each container. Fill each container with 250 ml of distilled 
water and stir the liquid until the powder has completely dissolved. 

5.3 

Electrode control  

Remove the protective cap from the electrode and check for damage.  
Damaged electrodes must be replaced. 
Rinse off any salt crystals that may have settled from the storage solution under running water.  
The salt crystals develop naturally and are not a deficiency. 
If the electrode has become dry, place it in a 3M KCL solution for at least 15 minutes. (storage 
solution for pH electrodes). 
 

 

Summary of Contents for PH 26F

Page 1: ...E PH 28L pH Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chan...

Page 2: ...tion 4 5 1 Insert batteries 4 5 2 Preparation of the buffer solution 4 5 3 Electrode control 4 5 4 Switch meter on off 5 5 5 Basic settings 5 6 pH calibration 6 6 1 1 point calibration 7 6 2 2 point c...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ement Measurement range 0 60 C 32 140 F Accuracy 1 C Resolution 1 C Calibration points 1 General specifications Temperature compensation 0 60 C 32 140 F Temperature of medium 0 60 C 32 140 F Environme...

Page 5: ...ter PCE PH 28L 1 x buffer 4 01 pH powder for 250 ml solution 1 x buffer 7 00 pH powder for 250 ml solution 1 x buffer 10 01 pH powder for 250 ml solution 3 x container for calibration solution 2 x bat...

Page 6: ...3 Align the guide groove with the electrode and carefully push the electrode into the meter Screw the union nut of the electrode tight 5 2 Preparation of the buffer solution Prepare 3 sealable contai...

Page 7: ...key being pressed The function can be enabled or disabled in the basic settings 5 5 Basic settings Press and hold the CAL key for 3 seconds to call up the basic settings Press the CAL key to call up t...

Page 8: ...ettings calibration is mandatory 6 pH calibration The devices allow a calibration of up to 3 points To achieve high measurement accuracy we recommend at least a 2 point calibration The meter automatic...

Page 9: ...hows pH 7 00 CAL1 or pH 6 86 CAL1 6 1 3 Immerse the electrode in the pH 7 00 or pH 6 86 buffer solution with the end of the electrode completely immersed in the calibration solution Gently stir the so...

Page 10: ...electrode in the pH 4 01 or pH 10 01 pH 9 18 buffer solution with the end of the electrode completely immersed in the calibration solution Carefully stir the solution with the electrode to obtain a ho...

Page 11: ...tely immersed in the calibration solution Stir the solution carefully with the electrode to obtain a homogeneous solution 6 4 4 Press the ENTER key The meter starts the calibration 6 3 5 After the mea...

Page 12: ...ration 7 Temperature calibration When deviating temperature values are displayed you can adjust them in the corresponding menu item 7 1 Press and hold the CAL key for 3 seconds to call up the basic se...

Page 13: ...he glass sensitive membrane moist Make sure that the electrode is thoroughly rinsed with distilled or deionised water after use If you do not use the meter for a long period of time store the electrod...

Page 14: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 15: ...hazlar Ltd ti Unit 11 Southpoint Business Park Via Pesciatina 878 B Interno 6 Halkal Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece...

Reviews: