© PCE Instruments
10
6.3.7.
press the "ENTER" key.
The meter starts the calibration.
6.3.8
After the measured value has stabilised,
the display shows the slope of the electrode in %
and then displays "End".
The 3-point calibration is completed.
Press the "MEAS" key to return to measuring mode even during calibration.
7
Temperature calibration
When deviating temperature values are displayed, you can adjust them in the corresponding
menu item.
7.1.
Press and hold the "CAL" key for 3 seconds
to call up the basic settings.
7.2.
Press the "Enter" key repeatedly until the
display shows "°C" and "CAL" or "°F" and "CAL"
.
7.3
Press the "CAL" key.
7.4
Press the "CAL" key again until the
temperature display matches the actual
temperature.
7.5
Press the "ENTER" key to confirm your entry.
7.6
Press the
"MEAS" key to return to measuring
mode.