5.4 

Switch meter on / off  

To switch on the meter, press and hold the "MEAS" key until the display switches on. 
Press and hold the "MEAS" key for 5 seconds to switch it off. 
The  auto-off  function  switches  the  unit  off  after  8  minutes  without  a  key  being  pressed.  The 
function can be enabled or disabled in the basic settings. 
 

 

5.5 

Basic settings  

Press and hold the "CAL" key for 3 seconds to call up the basic settings. 
Press the "CAL" key to call up the possible options of the adjustable items. 
Press the "ENTER" key to confirm the selection and jump to the next item. 
 

 

 
After confirming all selected items, the meter returns to measuring mode. 
By pressing the "MEAS" key, you can return directly to measuring mode. 
 

 

Summary of Contents for PH 26F

Page 1: ...E PH 28L pH Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chan...

Page 2: ...tion 4 5 1 Insert batteries 4 5 2 Preparation of the buffer solution 4 5 3 Electrode control 4 5 4 Switch meter on off 5 5 5 Basic settings 5 6 pH calibration 6 6 1 1 point calibration 7 6 2 2 point c...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ement Measurement range 0 60 C 32 140 F Accuracy 1 C Resolution 1 C Calibration points 1 General specifications Temperature compensation 0 60 C 32 140 F Temperature of medium 0 60 C 32 140 F Environme...

Page 5: ...ter PCE PH 28L 1 x buffer 4 01 pH powder for 250 ml solution 1 x buffer 7 00 pH powder for 250 ml solution 1 x buffer 10 01 pH powder for 250 ml solution 3 x container for calibration solution 2 x bat...

Page 6: ...3 Align the guide groove with the electrode and carefully push the electrode into the meter Screw the union nut of the electrode tight 5 2 Preparation of the buffer solution Prepare 3 sealable contai...

Page 7: ...key being pressed The function can be enabled or disabled in the basic settings 5 5 Basic settings Press and hold the CAL key for 3 seconds to call up the basic settings Press the CAL key to call up t...

Page 8: ...ettings calibration is mandatory 6 pH calibration The devices allow a calibration of up to 3 points To achieve high measurement accuracy we recommend at least a 2 point calibration The meter automatic...

Page 9: ...hows pH 7 00 CAL1 or pH 6 86 CAL1 6 1 3 Immerse the electrode in the pH 7 00 or pH 6 86 buffer solution with the end of the electrode completely immersed in the calibration solution Gently stir the so...

Page 10: ...electrode in the pH 4 01 or pH 10 01 pH 9 18 buffer solution with the end of the electrode completely immersed in the calibration solution Carefully stir the solution with the electrode to obtain a ho...

Page 11: ...tely immersed in the calibration solution Stir the solution carefully with the electrode to obtain a homogeneous solution 6 4 4 Press the ENTER key The meter starts the calibration 6 3 5 After the mea...

Page 12: ...ration 7 Temperature calibration When deviating temperature values are displayed you can adjust them in the corresponding menu item 7 1 Press and hold the CAL key for 3 seconds to call up the basic se...

Page 13: ...he glass sensitive membrane moist Make sure that the electrode is thoroughly rinsed with distilled or deionised water after use If you do not use the meter for a long period of time store the electrod...

Page 14: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 15: ...hazlar Ltd ti Unit 11 Southpoint Business Park Via Pesciatina 878 B Interno 6 Halkal Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece...

Reviews: