NMS044 Reference Manual Rev E
Setting System Alert Notification Details
4-8
Settings
When you change any setting or preference on the meter, the 831C
sends an alert notification.
Memory
The 831C sends an alert notification when the primary memory storage
(free space) is 25%, 10%, and 0%. Primary memory storage can be
internal or USB. For information about setting the primary memory
storage location, see
section 9.5
in the
831C SLM Reference Manual
.
Temperature
An alert notification will be sent when the internal temperature of the
meter is outside of 70°C and -40°C.
Power
There are 4 available alert notifications:
• Low external power: sent when power reaches External Shutoff
V 0.6 V
• Meter is shutting down due to low external battery: sent 4
seconds prior to shut down
• Power from mains (AC or solar power) is lost, and meter is
running on external battery for more than 30 minutes
• If power lost alert is sent, then power is restored to mains for at
least 3 minutes, an alert will be sent
Run State
This alert replaces the SLM State alert. It sends a notification for each
Run, Stop, Pause, or Reset on the meter.
Certification
You can send an alert notification during the meter’s startup if the
calibration date is expired, or if the reminder for calibration is due. This
alert occurs once every 24 hours, when enabled.
Clear Queue
The 831C often queues alert notifications to send at a later time if the
network connection is lost, etc. If you don’t want these older
notifications to send, click the
Clear Queue
button to delete them.