Model Number

699B02

HAND HELD SHAKER

Revision: C
ECN #: 49738

P erfo rm an ce

EN G L ISH

SI

Operating Frequency(± 1 %)

159.2 Hz

159.2 Hz

Acceleration Output(± 3 %)

1 g rms

9.81 m/s² rms

[1]

Velocity Output

0.39 in/sec rms

9.81 mm/s rms

[2]

Displacement Output

0.39 mil rms

9.81 µm rms

[2]

Transverse Output

≤ 3 %

≤ 3 %

Distortion(0 to 250 grams load)

7 %

7 %

[3]

Maximum Load

8.8 oz

250 gm

[4]

Automatic Switch Off Time

1.0 to 2.5 minutes

1.0 to 2.5 minutes

[5]

Calibration Cycles(250 gram load)

90 cycles

90 cycles

[6]

En viro n m en tal

Temperature Range(Operating)

15 to 130 °F

-10 to 55 °C

Electrical

Ramp-Up time

≤ 3 sec

≤ 3 sec

[7]

Power Required(Standard)

Internal Battery

Internal Battery

Power Required

(Alternate)

DC Power

DC Power

Internal Battery(Quantity)

4

4

Internal Battery

(Type)

AA

AA

[8]

DC Power(± 5 %)

12 VDC

12 VDC

DC Power

1.67 Amps

1.67 Amps

Battery Life(250 gram load)

2.3 hours

2.3 hours

[6]

P h ysical

Size - Height

7.8 in

200 mm

Weight(with batteries)

31 oz

900 gm

[7]

Mounting Thread

1/4-28 Female

No Metric Equivalent

[9][10]

Mounting Torque(Maximum)

to 10 in-lb

to 10 in-lb

[9]

Size - Diameter

2.2 in

56 mm

[11]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

 is a registered trademark of PCB Piezotronics, Inc.

OP TION A L  VERSION S

Optional versions have identical specifications and accessories as listed for the standard model

except where noted below. More than one option may be used.

N OTES:

[1]Unit supplied set to rms; see manual for peak selection.
[2]Calculated values for reference only.
[3]Typical max for range
[4]For sensors weighing <75 grams add model 080A136 mass
[5]Unit supplied set to auto shut off; see manual for continuous use selection.
[6]Approximate values, based on automatic switch off time and dependent on type of batteries.
[7]Typical.
[8]Alkaline type recommended for longest service life.
[9]Transducer to shaker table.
[10]Test sensor should be hand tightened (without tools).
[11]See PCB Declaration of Conformance PS022 for details.

Date: 07/10/2019

Date: 07/10/2019

Date: 07/10/2019

Date: 07/10/2019

67896

Entered: LK

Engineer: BAM

Sales: MC

Approved: BAM

Spec Number:

P h o n e: 800-959-4464
F ax: 716-684-3823
E-M ail: im i@ p cb .co m

3425 Walden Avenue, Depew, NY 14043

SU P P L IED A C C ESSORIES:

Model M081A19 Mounting stud, 1/4-28 to M6 x 1, SS with shoulder (1)
Model 073A15 Battery Pack (1)
Model 080A136 Calibration Mass (1)
Model 081A08 Mounting Stud (10-32 to 1/4-28) (1)
Model 081A40 Mounting Stud (2)

Summary of Contents for IMI DIVISION 699B02

Page 1: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 2: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 3: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 4: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 5: ...N FOR MACHINE CONDITION MONITORING Model 699B02 Handheld Vibration Shaker Operating Guide with Enclosed Warranty Information 3425 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684...

Page 6: ...NSORS AND INSTRUMENTATION FOR MACHINE CONDITION MONITORING Table of Contents Introduction page 3 Description page 3 Maintenance Repair page 6 ESD Warning Information page 7 Warranty Servicing Warranty...

Page 7: ...ker Description Model 699B02 permits accurate adjustment of measuring instrumentation to indicate a standard acceleration level of 1 g RMS or 1 g peak adjusting the system for correct measurement The...

Page 8: ...indicating instrument may now be adjusted to read the relevant reference value Following the system adjustment the shaker is switched off by pressing the On Off button a second time To prolong the us...

Page 9: ...lly disconnected Figure 3 Peak vs RMS Selection with 100 mV g Accelerometer Calibration Check Points Calibration of the internal feedback circuit can be verified via test points located in the options...

Page 10: ...If factory service is required return the instrument to IMI A free quotation will be provided prior to servicing IMI a division of PCB Piezotronics Inc is an ISO 9001 certified company that has embra...

Page 11: ...voltage electrostatic discharge ESD can damage electrical devices Similar to a capacitor a cable can hold a charge caused by triboelectric transfer such as that which occurs in the following Laying on...

Page 12: ...ion will be provided prior to servicing at no charge Before returning the unit please consult a factory IMI applications engineer concerning the situation as certain problems can often be corrected wi...

Page 13: ...o have your purchase price refunded IMI offers to all customers at no charge 24 hour phone support This service makes product or application support available to our customers day or night seven days...

Page 14: ...d In the interest of constant product improvement we reserve the right tochange specifications without notice ICP is a registeredtrademarkof PCB Piezotronics Inc OPTIONAL VERSIONS Optional versions ha...

Page 15: ...II 003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 8792 1 OF 1 625X MODEL 699A02 699B02 HAND HELD CALIBRATOR DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 11 1 18 KRM 11 1 18 CPH 11 1 18 8792...

Reviews: