PBI SAS SUPER IAQ Instruction Manual Download Page 25

 

                                           “

SAS  Super IAQ”                   Rev. 1                                 Pg. 

21

 

ENGLISH 

CORRECTION TABLE TO ADJUST COLONY COUNTS FROM A 401-HOLE 

IMPACTOR USING STANDARD 90 mm PETRI PLATES  

Colon

ies 

MP

Pr 

Coloni

es 

MP

Pr 

Coloni

es 

MP

Pr 

Coloni

es 

MP

Pr 

Coloni

es 

MP

Pr 

Colo

nies 

MP

Pr 

1 1 41 43 81 90 

121 

144 

161 

206 

201 

278 

2 2 42 44 82 92 

122 

145 

162 

207 

202 

280 

3 3 43 45 83 93 

123 

147 

163 

209 

203 

282 

4 4 44 47 84 94 

124 

148 

164 

211 

204 

284 

5 5 45 48 85 95 

125 

150 

165 

212 

205 

287 

6 6 46 49 86 97 

126 

151 

166 

214 

206 

289 

7 7 47 50 87 98 

127 

152 

167 

216 

207 

291 

8 8 48 51 88 99 

128 

154 

168 

217 

208 

293 

9  9  49  52  89  100 129 155 169 219 209 295 

10 10 50 53 90 102 

130 

157 

170 

221 

210 

297 

11 11 51 54 91 103 

131 

158 

171 

223 

211 

299 

12 12 52 56 92 104 

132 

160 

172 

224 

212 

301 

13 13 53 57 93 106 

133 

161 

173 

226 

213 

303 

14 14 54 58 94 107 

134 

163 

174 

228 

214 

305 

15 15 55 59 95 108 

135 

164 

175 

230 

215 

307 

16 16 56 60 96 110 

136 

166 

176 

231 

216 

310 

17 17 57 61 97 111 

137 

167 

177 

233 

217 

312 

18 18 58 63 98 112 

138 

169 

178 

235 

218 

314 

19 19 59 64 99 114 

139 

170 

179 

237 

219 

316 

20  20  60  65  100 115 140 172 180 239 220 318 
21  22  61  66  101 116 141 173 181 240 221 321 
22  23  62  67  102 118 142 175 182 242 222 323 
23  24  63  68  103 119 143 177 183 244 223 325 
24  25  64  70  104 120 144 178 184 246 224 327 
25  26  65  71  105 122 145 180 185 248 225 330 
26  27  66  72  106 123 146 181 186 250 226 332 
27  28  67  73  107 124 147 183 187 251 227 334 
28  29  68  74  108 126 148 184 188 253 228 336 
29  30  69  76  109 127 149 186 189 255 229 339 
30  31  70  77  110 128 150 188 190 257 230 341 
31  32  71  78  111 130 151 189 191 259 231 343 
32  33  72  79  112 131 152 191 192 261 232 346 
33  34  73  80  113 133 153 192 193 263 233 348 
34  35  74  82  114 134 154 194 194 265 234 351 
35  37  75  83  115 135 155 196 195 267 235 353 
36  38  76  84  116 137 156 197 196 269 236 355 
37  39  77  85  117 138 157 199 197 271 237 358 
38  40  78  87  118 140 158 201 198 272 238 360 
39  41  79  88  119 141 159 202 199 274 239 363 
40  42  80  89  120 142 160 204 200 276 240 365 

 

 

 

Summary of Contents for SAS SUPER IAQ

Page 1: ...ine on Sterile Drug Products produced by Aseptic Process ACGIH Guideline for the Assessment of Bioaerosol in the Indoor Environment ASTM Draft Protocol Committee D22 05 06 USP 23 NF 18 8th Supplement...

Page 2: ...ays disconnect the charger before A Repairing or maintenance these operations must be carried out by qualified staff B Keep the unit clean Use original spare parts and accessories for any replacement...

Page 3: ...te holder adjusting 6 3 5 84 mm Maxi Contact Plate use optional 6 3 6 Tripod Installation optional 7 4 Operations 7 4 1 Start with the previous volume 8 4 2 User Mode 8 4 3 Program mode 9 4 4 Utility...

Page 4: ...itres 19 SECTION THREE 7 Specifications 23 7 1 Messages Shown On Display 25 7 2 Sampling Time Requested For Different Air Volumes 26 7 3 Battery Charger Operation 26 7 4 Automatic Switch OFF 27 SECTIO...

Page 5: ...reach the sub menu of the chosen parameter Refer to instruction manual to change settings Press ON OFF switch at the end of sampling operations Press CLEAR each time you need to end an action Then th...

Page 6: ...the Surface Air System SAS are A To use a simple and inexpensive Contact Plate SURFAIR PLATE RODAC for surface hands or air control These plates are very well known and easily available They can be pu...

Page 7: ...closed and prepared Contact Plate and remove plate lid 3 Replace the aspirating head 4 Select the required air volume and start the unit The air flow is directed onto the agar surface of the Contact...

Page 8: ...eight programmable sampling volumes DELAY MODE With this function the air sampler can be set to start after a programmable delay UTILITY MODE This function led to the following Sub Menu SET TIME To a...

Page 9: ...cedures should be followed to check that the unit is working properly The battery pack of the instrument must be charged for at least 4 hours before the test 14 hours if slow battery charger is used S...

Page 10: ...CONTACT PLATES HOLDER ADJUSMENT The Contact Plate holders can be adjusted using a screw driver in case the available plates slightly different in diameter from the standard 55 mm Contact Plate see th...

Page 11: ...he unit if the automatic switch off timer operated when last used Press CLEAR at any time to return to the initial START FOR configuration The air sampler is provided with eight programmable air volum...

Page 12: ...USER MODE PROGRAM Up to eight volumes can be programmed into the unit ACTION DISPLAY Switch on the ON OFF switch S T A R T F O R X X black button M D Y H M Press to select U S E R M O D E the USER MOD...

Page 13: ...D Y H M Press to select P R O G R A M M O D E PROGRAM MODE M D Y H M Press ENTER and then S P R O G X X X X to select the volume of air to be modified M D Y H M Press ENTER X X X X 0 0 0 0 M D Y H M P...

Page 14: ...press U T I L I T Y M O D E to reach UTILITY MODE M D Y H M 4 4 1 SET TIME This option is used to program day month year and time of the day ACTION DISPLAY Press ENTER to select S E T T I M E the SET...

Page 15: ...nd 3rd and 4th letter or number X X X X S A M P L I N G S I T E X X X X S A M P L I N G S I T E X X X X Press ENTER to confirm 4 4 3 IDENTIFY This option is used to identify the operator This should b...

Page 16: ...be selected choosing from different languages Select UTILITY MODE Program ACTION DISPLAY Press ENTER and then to select the L A N G U A G E LANGUAGE Sub Menu M D Y H M Press ENTER and than to select...

Page 17: ...UTILITY MODE Program ACTION DISPLAY Press ENTER and then to select the D I S P L A Y R E C O R D DISPLAY RECORD Sub Menu M D Y H M Press ENTER and then to select the number on the left X X M D Y H M s...

Page 18: ...OGRAM TO DELAY INSTRUMENT START ACTION DISPLAY Switch on the ON OFF S T A R T F O R X X switch black button M D Y H M Press to select delay D E L A Y M O D E mode M D Y H M Press ENTER to reach the de...

Page 19: ...utton 4 7 REMOTE SWITCH USE The SAS SUPER IAQ microbiological air samplers can be fitted with an Infrared Remote Control optional The instrument in normal operating conditions automatically switches o...

Page 20: ...now not switch off after 4 minutes and the Infrared Remote Switch will work regularly ACTION DISPLAY Press clear to go S T A R T F O R X X back to the start D M Y H M NOTE At the end of this protocol...

Page 21: ...t is therefore useful to have available several heads ready to use Alternatively the unit can be sanitised using a disinfecting aerosol at about 30 cm from the head when the fan is running Thirty seco...

Page 22: ...e air path C BODY OF THE UNIT The plastic body of the unit can be wiped with a mild chemical agent Note The air sampler is a scientific instrument and its surface should be disinfected using an aeroso...

Page 23: ...minute The usual plating technique and or identification procedures are applied to the processed diluent 5 FINAL OPERATIONS Switch off the ON OFF Main switch black button The unit will switch off auto...

Page 24: ...64 301 201 542 15 15 52 59 89 114 128 192 165 305 202 554 16 17 53 60 90 116 129 194 166 309 203 567 17 18 54 62 91 117 130 196 167 313 204 580 18 19 55 63 92 119 131 199 168 317 205 595 19 20 56 64 9...

Page 25: ...96 110 136 166 176 231 216 310 17 17 57 61 97 111 137 167 177 233 217 312 18 18 58 63 98 112 138 169 178 235 218 314 19 19 59 64 99 114 139 170 179 237 219 316 20 20 60 65 100 115 140 172 180 239 220...

Page 26: ...of air r Colony Forming Units counted on 55 mm Contact Plates 67 Pr Probable count obtained by positive hole correction 80 x Colony Forming Units per 1000 litres 1 cubic metre of air 80 x 1000 x 400...

Page 27: ...sampling efficiency of the SAS SUPER IAQ in a controlled environment with aerosol of known particle size is 100 over 4 microns in size Considering the normal range of environmental micro organisms is...

Page 28: ...er connection 16 Flashing LED during air sampling 17 Instrument Serial Number 18 CE mark SAS Super IAQ Front Control Panel 19 Infrared sensor for remote control 20 Screw for battery pack and access to...

Page 29: ...options S SAS IAQ Name of instrument ID XXXX Operator s identification SITE XXXX Sampling location identification START FOR XX The instrument is ready to collect the same volume of air as the last sam...

Page 30: ...hours using the normal battery charger Revit Cat n 19248 or after 3 5 hours using the fast battery charger Cat n 88603 5 Disconnect the battery charger from the mains supply 6 At the end of operation...

Page 31: ...per day To prevent injury or burns do not allow metal objects to contact or short circuit the battery terminals The Metal Hydrate battery may explode if disposed in fire 7 4 Automatic switch off The i...

Page 32: ...uld be validated at least every six twelve months The validation of the air flow rate should be done by a certified anemometer in a wind tunnel This type of certification may be performed by PBI inter...

Page 33: ...place infrared remote battery Microbiological media is dehydrated after sampling Control expiration date of media and that the agar is not dehydrated before sampling 8 5 Battery replacement 1 Take out...

Page 34: ...t n 19248 Normal battery charger 110 Volt Cat n 86603 Fast battery charger 220 Volt Cat n 20153 Aluminium carrying case Cat n 18143 Soft carrying case Cat n 18950 Infrared remote control Cat n 18200 I...

Page 35: ...after the visual display presentation D M Y H M TO START WITH ONE OF THE EIGHT FREE SAMPLING VOLUMES PRE MEMORISED IN THE USER MODE PROGRAM ACTION DISPLAY Switch on the ON OFF switch S T A R T F O R...

Page 36: ...A 60 With Plate Count Agar for total viable count 80 7426 PCA TW 60 With Plate Count Agar and neutralizing agent for total viable count 80 17354 TSA TW 60 With Tryptone Soy Agar and neutralizing agent...

Page 37: ...n XXXX Date XXXX Company Institute XXXX Laboratory XXXX Name of Instrument Microbiological Air Sampler Model SAS SUPER IAQ Serial Number XXXX Name of Producer International PBI Name of Distributor XX...

Page 38: ...ASKED TO THE EXPORT DPT Company Address City Zip Code State Phone Fax Email Name to contact Lab Division return request of technical service Equipment Model Serial Number for calibration repair other...

Page 39: ...SAS Super IAQ Rev 1 Pg 35 ENGLISH...

Page 40: ...ges resulting from tamperings abnormal use and installation and does not cover transport damages International PBI shall not be liable for any commercial damages whether incidental consequential or ot...

Page 41: ...sole responsibility that the following product SAS Super IAQ code 90578 the following harmonized standards have been applied EN 61010 1 2001 EN 61326 1 2000 and conforms with the technical requiremen...

Reviews: