10. Specifications
MAP Check 9000 - User Manual
40
10.4. Flow system MAP Check PFC
MAP Check PFC also contains a system, which measures out the flushing gas using input
from the oxygen sensor. This keeps the oxygen level within the desired limits.
The sensor flow system is like a standard MAP Check.
Fig. 7. MAP Check PFC flow diagram
10.4.1. Specifications PFC
•Regulation gas:
Up to 1000L/min (STP)
•Max input pressure:
6.5bar (Stationary)
•Min input pressure:
1bar
•Connection:
Hose connection, 260M-12-1/2 (PBI 930505)
O
sensor
2
Flow
sensor
Proportional
valve
Pressure
sensor
Thrush
valve
Approx. 200 ml/min
0.2 - 1.0
l/min
Pump
Gas In
Regulated Gas In
Gas Out
Regulated Gas Out