![PBI Dansensor MAP Check 9000 User Manual Download Page 10](http://html1.mh-extra.com/html/pbi-dansensor/map-check-9000/map-check-9000_user-manual_1513964010.webp)
PBI-Dansensor A/S
COPYRIGHT ©
PBI-220047-F 01/2008
9
MAP Check 9000 - User Manual
3. Connection
3. Connection
3.1. MAP Check 900X-1/2/4
•Pump control signal for external pump control (10-32VDC)
•Alarm relay outputs
•System error / Ready relay output
•Current/voltage outputs (optional)
•CAN bus (optional)
Fan / dust filter
Mains voltage
Mains fuse
RS232C port for serial communication with PC, etc.
ST1
ST2
Gas in
Gas out
Rating plate
ST1 1,2: Pump control signal, 10-32VDC input (for external pump control)
3,4: O
2
alarm 1 relay. Max. 48V, 1A
5,6: O
2
alarm 2 relay. Max. 48V, 1A
7,8: System error / Ready relay. Max. 48V, 1A
ST2 1,2: * Reserved for current output (optional)
(1+, 2-)
3,4: * Reserved for CAN bus (optional)
(3+, 4-)
5,6
* Reserved for 24VDC output (optional)
(5+, 6-)
* Other connection will be marked with a text on the back of MAP Check 9000
3
2
4
7
8
5
6
1
9