Page 29
SySt.FS
CAPACITY OF THE WEIGHING SYSTEM [1301-1302]
Programming the useful capacity (net) of the weighing system.
Values: from 0 to Load Cell Capacity
Default: 0
dEAd L.
FIXED TARE OF WEIGHING SYSTEM [1106-1107]
Programming the fixed tare value of the weighing system
Values: from 0 to Capacity Value
Default: 00000
CAL.tyP
SELECTING THE CALIBRATION TYPE
Selecting the calibration type. On confirmation one of the following procedures is started.
dEAd W
DEAD WEIGHT TYPE CALIBRATION [501÷503]
Zero Calibration and Full Scale up to 5 linearisation points using Sample weights.
tAbLE
TABLE TYPE CALIBRATION [1151÷1172]
Allows you to manually program up to 5 calibration points. The values corresponding to those resulting
from the linearisation procedure with sample weights. In this way you can copy the calibration values
made with sample weights.
PARAMETERS DISPLAYED ONLY DURING METRIC OPERATION
G-CAL
CALIBRATION LOCATION GRAVITY [1108-1109]
Programming of the force of gravity of the location where the calibration takes place.
Values: from 9.77000 to 9.84000
Default: 0
G-uSE
GRAVITY OF LOCATION OF USE [1110-1111]
Programming of the force of gravity of the location where the instrument will be used.
Values: from 9.77000 to 9.840000
Default: 0