Page 25 Instruction manual Parker X-Flow FM-1407 Rev - 11/2021
Green
LED
Time
Indication
Off
Continuously Power-off or program not running
On
Continuously Normal running/operation mode
Flash
0.2 sec on,
0.2 sec off
Special function mode
Instrument is busy performing any special function.
E.g. auto-zero or self-test
Green LED indication modes (no switch used)
Red LED
Time
Indication
Off
Continuously No error
Flash
Variable
Bus activity on the Modbus® interface
On
Continuously Critical error message
A serious error occurred in the instrument
Instrument needs service before further using
Red LED indication modes (no switch used)
3.9 Basic Parameters and Properties
3.9.1 Introduction
Every parameter has its own properties. These properties are given in a table as shown:
Type
Access Range FlowDDE Flowbus® Modbus®
[type]
RW
[x]…[y] [FB]
[Pro]/[Par] [address]/[index]
Type
Unsigned char 1-byte character
Unsigned char[x] x-byte array (string)
Unsigned int
2-byte unsigned integer
Float
4-byte floating point
Access
R
The parameter is read-only
RW
The parameter can be read and write
RW
The parameter can only be written when the Init Reset parameter is set to 64. See section
4.1.1, General Product Information for more details.
Range
Some parameters only accept values within a certain range:
[x]
Minimal value of the range.
[y]
Maximal value of the range.
FlowDDE
Parameter number in FlowDDE. Check section 3.2, “In analog operation following signals are available:
Measured value (analog output)
Setpoints below 2% of the full scale will be interpreted as 0% setpoint.
When operating the instrument through the analog interface it is possible to connect the
instrument simultaneously to RS232 or Modbus®/RS485 for reading/changing parameters